Development and clinical validation of a clinically translatable non-chip-on-tip transvaginal imaging system (GynoSight v2.0) for early detection of premalignant cervical lesions.

Jeyachandran Karthika, Pakirithodiyil Mohammed Ansar, Aruldoss Keerthana, Lal Milind, M Dhanush Koodi, Anantharaju Arpitha, Sun YuanYuan, Liang Rongguang, Pal Uttam M

СтатьяВалидационное исследованиеПеревод в очереди

Аннотация

SIGNIFICANCE: Cervical cancer is the fourth most common cancer among women globally. Hence, it is crucial to develop a noninvasive and portable optical imaging modality for the early detection of premalignant cervical lesions.

AIM: We present the development and clinical validation of GynoSight v2.0, an indigenously developed multispectral, non-chip-on-tip source, hand-held, portable transvaginal imaging probe, for evaluating tissue health and identifying anomalies, such as those linked to precancerous cervical lesions.

APPROACH: GynoSight v2.0 houses a 16 LEDs, 5-megapixel camera, and a Raspberry Pi 5 module. A comparative shadowing effect analysis was performed between GynoSight v2.0 and colposcopy by evaluating statistical metrics such as mean pixel intensity (MPI), shadow area percentage (SAP), entropy, and contrast-to-noise ratio. In addition, the relative oxygen saturation maps of the cervical tissue were computed from the multispectral registered image using the proposed discrete Fourier transform-based image registration technique.

RESULTS

The images of N = 6 ( N = 2 normal, N = 1 premalignant, and N = 3 high-grade squamous intraepithelial lesion) subjects were acquired. A comparative shadowing analysis shows a 50-unit gray level value separation between the colposcope and GynoSight v2.0 images. The pixel values of the colposcope are skewed to the lower pixel values, and the pixel values of GynoSight v2.0 are spread uniformly over 0 to 255 gray-level pixel values. In addition, the statistical analysis showed that MPI, SAP, and entropy are significant metrics for shadowing effect quantification.

CONCLUSIONS

The colposcopy images showed more shadowing effects than the GynoSight v2.0 images and hence provide better illumination to aid in better diagnosis.

Для цитирования: Jeyachandran Karthika, Pakirithodiyil Mohammed Ansar, Aruldoss Keerthana, Lal Milind, M Dhanush Koodi, Anantharaju Arpitha, Sun YuanYuan, Liang Rongguang, Pal Uttam M. Development and clinical validation of a clinically translatable non-chip-on-tip transvaginal imaging system (GynoSight v2.0) for early detection of premalignant cervical lesions.. Journal of biomedical optics. 2025. DOI: 10.1117/1.jbo.30.10.106002.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.