DOI 10.1055/a-2889-7878

Медицинская библиотека · Veterinary and comparative orthopaedics and traumatology : V.C.O.T · 2026

Computed Tomographic Measurement of Cementless Total Hip Replacement Implant Orientation.

Pink Jonathan, Bilmont Alexis

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

The aim of this study was to describe and assess a technique for the measurement of cementless (BioMedtrix Biologic Fixation [BFX]) total hip replacement implant orientation using computed tomography (CT).

STUDY

DESIGN

Postoperative CT studies of 15 dogs that had total hip replacement with a 24 mm BFX cup and BFX stem were reviewed retrospectively. Multiplanar reconstructed CT images were manipulated to generate standard positions of the pelvis and femur. Open face version, angle of lateral opening (ALO), inclination of the cup and stem anteversion were measured sequentially. Cases were anonymized, randomized and measured three times by two observers.

RESULTS

Intraclass correlation coefficients for cup open face version, cup ALO, cup inclination and stem version were 1.0 and 1.0, 0.98 and 0.98, 0.99 and 0.99, and 1.0 and 0.97, for observers 1 and 2, respectively. The interclass correlation coefficients indicating interobserver variation for cup open face version, cup ALO, cup inclination and stem version were 1.0, 0.99, 0.99 and 0.99, respectively.

CONCLUSION

The reported technique for measurement of BFX total hip replacement implant orientation is highly repeatable and allows objective measurement of implant orientation, including parameters which are not accessible radiographically.

Для цитирования: Pink Jonathan, Bilmont Alexis. Computed Tomographic Measurement of Cementless Total Hip Replacement Implant Orientation.. Veterinary and comparative orthopaedics and traumatology : V.C.O.T. 2026. DOI: 10.1055/a-2889-7878.