Comparison between Air-Q Self Pressurized Airway Device with Blocker and Proseal Laryngeal Mask Airway in anesthetized paralyzed adult female patients undergoing elective gynecological operations.

Ismail Youssef Maha Mohammed, Dobal Naser Mohammed, Hammad Yahya Mohamed, El-Refai Nesrine Abdel Rahman, Abdelrahman Reham Ali Abdelhaleem

СтатьяРандомизированное исследованиеСравнительное исследованиеПеревод в очереди

Аннотация

INTRODUCTION

The Air-Q Self Pressurized Airway Device with Blocker (SP Blocker) was compared to the Proseal Laryngeal Mask Airway (PLMA) during positive pressure ventilation regarding the primary outcome (oropharyngeal leak pressure [OLP]), secondary outcomes (peak inspiratory pressure [PIP], inspired tidal volume [ITV], expired tidal volume [ETV], leak volume [LV] and leak fraction [LF]), insertion time, ventilation score, fiber-optic glottis view score, and postoperative laryngopharyngeal parameters (LPM).

MATERIAL AND

METHODS

Adult healthy female patients scheduled for elective gynecological laparotomies under general anesthesia using controlled mechanical ventilation were recruited to a prospective randomized comparative clinical trial. Exclusion criteria were body mass index (BMI) ≥ 35 kg m -2 , El-Ganzouri score ≥ 5, upper airway problems, hiatus hernia or pregnancy. Patients were classified into an SP Blocker group ( n = 75) and a PLMA group ( n = 75). Primary and secondary outcomes were assessed initially and at fixed time points after successful insertion of devices.

RESULTS

Initially after successful device insertion: the SP Blocker group showed statistically significant higher mean OLP (cmH 2 O) (29.46 ± 2.11 vs. 28.06 ± 1.83 respectively; 95% CI: -2.037 to -0.76, P < 0.0001), lower mean PIP (cmH 2 O) (15.49 ± 0.61 vs. 17.78 ± 1.04 respectively; 95% CI: 2.02 to 2.56, P < 0.0001), higher mean ITV (mL) (411 ± 30 vs. 403 ± 15 respectively; 95% CI: -15.65 to -0.347, P = 0.041), higher mean ETV (mL) (389 ± 12 vs. 354 ± 11 respectively; 95% CI: -38.72 to -31.29, P < 0.0001), lower mean LV (mL) (22 ± 18 vs. 49 ± 10 respectively; 95% CI: 22.3 to 31.7, P < 0.0001) and lower mean LF (%) (5 ± 2.04 vs. 12 ± 6.8 respectively; 95% CI: 5.38 to 8.62, P < 0.0001) than the PLMA group. Mean insertion time (seconds) was shorter in the SP Blocker group than the PLMA group (16.39 ± 2.81 vs. 18.63 ± 3.44 respectively; 95% CI: 1.23 to 3.25, P < 0.0001). The SP Blocker group offered a better fiber-optic glottis view score than the PLMA group without differences concerning ventilation score and LPM.

CONCLUSIONS

SP Blocker provided as safe anesthesia during controlled mechanical ventilation as PLMA.

Для цитирования: Ismail Youssef Maha Mohammed, Dobal Naser Mohammed, Hammad Yahya Mohamed, El-Refai Nesrine Abdel Rahman, Abdelrahman Reham Ali Abdelhaleem. Comparison between Air-Q Self Pressurized Airway Device with Blocker and Proseal Laryngeal Mask Airway in anesthetized paralyzed adult female patients undergoing elective gynecological operations.. Anaesthesiology intensive therapy. 2024. DOI: 10.5114/ait.2024.141203.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.