BACKGROUND

We evaluated whether intraoperative neural monitoring (IONM), including somatosensory evoked potential monitoring (SEP), motor evoked potential monitoring (MEP), and electrophysiological monitoring (EMG), could reduce operative errors attributed to lumbar instrumented fusions, including minimally invasive (MI) transforaminal lumbar interbody fusion (TLIF)/open TLIF.

METHODS

Operative errors included retraction/stretch or cauda equina neural/cauda equina injuries that typically occurred during misplacement of interbody devices (IBD) and/or malpositioning of pedicle screws (PS).

RESULTS

IONM decreased the incidence of intraoperative errors occurring during instrumented lumbar fusions (MI-TLIF/TLIF). In one series, significant loss of intraoperative SEP in 5 (4.3%) of 115 patients occurred after placing IBD; immediate removal of all IBD left just 2 patients with new neural deficits. In other series, firing of trigger EMG's (t-EMG) detected intraoperative PS malpositioning, prompted the immediate redirection of these screws, and reduced the need for reoperations. One t-EMG study required a reoperation in just 1 of 296 patients, while 6 reoperations were warranted out of 222 unmonitored patients. In another series, t-EMG reduced the pedicle screw breech rate to 7.78% (1723 PS) from a higher 11.25% for 1680 PS placed without t-EMG. A further study confirmed that MEP's picked up new motor deficits in 5 of 275 TLIF.

CONCLUSION

SEP/MEP/EMG intraoperative monitoring appears to reduce the risk of surgical errors when placing interbody devices and PS during the performance of lumbar instrumented fusions (MI-TLIF/TLIF).However, IONM is only effective if spine surgeons use it, and immediately address significant intraoperative changes.

Для цитирования: Epstein Nancy E. Perspective: Can intraoperative neurophysiological monitoring (IONM) limit errors associated with lumbar pedicle screw fusions/transforaminal lumbar interbody fusions (TLIF)?. Surgical neurology international. 2023. DOI: 10.25259/sni_671_2023.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.