DOI 10.1007/s00270-026-04545-w
Robotic-Assisted Computed Tomography-Guided Percutaneous Bone Procedures: A Prospective Multicenter Pilot Study.
Аннотация
PURPOSE: To prospectively evaluate the feasibility, safety, and targeting accuracy of robotic-assisted CT-guided percutaneous bone procedures.
MATERIALS AND
METHODS
This prospective multicenter pilot study included adult patients undergoing CT-guided percutaneous bone consolidation or tumor ablation between September 2024 and January 2025 at three academic institutions. All procedures were planned and performed using the Epione® robotic platform. Targets included the thoracic and lumbar spine and the pelvis. The primary endpoint was feasibility of robotic-assisted needle insertion, defined as successful needle placement without conversion to manual technique. Secondary endpoints included safety, targeting accuracy, procedural time, and radiation dose. Targeting accuracy was measured as lateral tip deviation and angular deviation between planned and final needle positions.
RESULTS
Thirty-four patients underwent 90 needle insertions, including 66 in the spine (73.3%) and 24 in the pelvis (26.7%). Indications were bone consolidation in 87/90 insertions (97%) and tumor ablation in 3/90 (3%). Robotic insertion was feasible in 89/90 insertions (98.9%). Seven adverse events occurred in 6 patients (17.6%), including four serious events, none related to the device or procedure. Mean lateral tip deviation was 2.0 ± 1.4 mm, and mean angular deviation was 3.1° ± 2.7°. Mean dose-length product was 3224 ± 2215 mGy.cm. Mean total procedure time was 03 h 09 min ± 50 min.
CONCLUSIONS
Robotic-assisted CT-guided bone procedures showed high feasibility and satisfactory targeting accuracy, with no device- or procedure-related adverse events. These findings support further evaluation in larger comparative studies.
CLINICAL TRIAL REGISTRATION: NCT06376682.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.