DOI 10.1245/s10434-023-14552-7
Feasibility of Autofluorescence Using Overlay Imaging for the Detection of Parathyroid Glands: Defining Standards.
Аннотация
BACKGROUND
The aim of this study is to define standards for the use of near-infrared autofluorescence (NIRAF)-based overlay imaging via EleVision IR (Medtronic, Dublin, Ireland) and to evaluate its clinical applicability.
PATIENTS AND
METHODS
This prospective study included 189 patients who had undergone open thyroid and/or parathyroid surgery and in whom EleVision IR was applied to visualize at least one parathyroid gland (PG) between January 2021 and May 2022 in a tertiary referral care center. Whether the PGs were first localized by the surgeon or by overlay imaging was noted. Handling of the device, application time and duration, distance, infrared intensity (IR%), and the angle of each measurement were analyzed. In thyroidectomies, the specimens were subsequently scanned for further PGs. NIRAF patterns and intensities were described.
RESULTS
Overall, 543 PGs were analyzed in 158 (83.6%) surgeries of thyroid glands (TGs) and in 49 (25.9%) surgeries for hyperparathyroidism. In 111 (58.7%) patients, identical numbers of PGs were detected by the surgeon and by overlay imaging. While a larger number of PGs was identified by the surgeon in 48 (25.4%) patients, overlay imaging served to detect more PGs in 30 (15.9%) cases. In four (2.1%) patients, PGs were visualized post-thyroidectomy due to their autofluorescence on the specimen. NIRAF-based overlay imaging was applied to depict the PGs early on after exposure by the surgeon. The ideal distance for the measurement ranged between 8 and 12 cm with an angle of 90° and a mean IR% of 34.5% (± 17.6).
CONCLUSIONS
Considering the standard operating procedures, NIRAF-based overlay imaging can be used as an adjunct tool for intraoperative localization.