Estimating Range of Vision and Optical Function Related to Pupil Size in a Spiral Intraocular Lens.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Kim Dong Hyun, Son Hyeck-Soo, Wu Zhiyi, Auffarth Gerd U, Łabuz Grzegorz

Аннотация

PURPOSE: To perform a comprehensive in vitro optical assessment of the RayOne Galaxy intraocular lens (IOL) (Rayner Intraocular Lenses, Ltd) to characterize its optical quality and functional visual range.

METHODS: Three samples of the +20.00 diopters (D) RayOne Galaxy IOL were assessed using the OptiSpheric IOL PRO2 (Trioptics GmbH) bench. All measurements used polychromatic light and a model cornea inducing +0.27 µm of spherical aberration. Modulation transfer function (MTF), through-focus MTF (TF-MTF), area under the MTF (MTFa), and simulated visual acuity (SimVA) were evaluated across pupil diameters of 2 to 4.5 mm. Photic phenomena susceptibility was also assessed via point spread function (PSF) imaging.

RESULTS: The Galaxy demonstrated robust optical quality and an extended functional vision for small pupils (2 to 3 mm). At these apertures, the SimVA remained ≤0.2 logarithm of the minimum angle of resolution (logMAR) across a wide defocus range (+1.00 to -3.00 D). The Galaxy showed a distinct higher plateau in the 0.00 to -2.00 D defocus range on the TF-MTF curve for small pupils. PSF analysis also demonstrated a compact central peak and minimal halo rings. However, a significant decline in objective metrics, including SimVA, was observed with increasing defocus for pupils >3.5 mm.

CONCLUSIONS: The Galaxy demonstrated an extended range of vision with good optical quality and a low degree of positive dysphotopsia, particularly at small pupil sizes. Optical performance at intermediate and near distances tends to decline notably as pupil diameter increases.