MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3389/fonc.2022.872890

Медицинская библиотека · Frontiers in oncology · 2022

Comparison of Contrast-Enhanced Ultrasonography to Color Doppler Ultrasound in Evaluation of Carotid Body Tumors.

Gu Guangchao, Zhang Xiaoyan, Shen Junyue, Gulidanna Shayan, Gao Qiong, Shao Jiang, Liu Bao, Zhang Bo, Zheng Yuehong

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

The objectives of this study were to prospectively 1) explore the characteristics and enhanced patterns of carotid body tumors (CBTs) at color Doppler ultrasound (CDU) and contrast-enhanced ultrasonography (CEUS) qualitatively and quantitatively and 2) compare CDU and CEUS for their morphology and vascularity signature.

METHODS

CDU and CEUS with Sonovue® were used to evaluate 25 CBT lesions. The comparison between these ultrasonic modalities included the size, Shamblin type, vascularity, and feeding vessels of the lesion areas. The time-intensity curve (TIC) analysis was used to obtain the dynamics of the contrast-enhancement features of CBTs.

RESULTS

The TIC analysis presented a fast wash-in [wash-in time: 3.00 ± 1.10 s, mean ± SD] and slow wash-out [wash-out time: 58.79 ± 24.21 s, mean ± SD] pattern in the CBT lesions, with a high area under the curve (AUC) of 669.68 ± 143.46 mm2 (mean ± SD). In comparison with CDU, CEUS was superior in identifying Shamblin type I or III CBT lesions (χ2 = 17.389, p=0.002). It detected a significant difference in the AUC between moderate and marked vascularity groups (563.33 ± 102.63 vs. 707.22 ± 138.81, t=-2.311, p=0.031.), while CDU observed no significant difference between these two groups. Although CDU was more sensitive than CEUS in detecting feeding vessels (100% vs. 88%), CEUS better visualized the origins of feeding vessels (χ2 = 9.162, p=0.010).

CONCLUSION

CEUS can better investigate the Shamblin type and vascularity of CBT lesions than CDU. CBTs displayed a fast wash-in, slow wash-out pattern with high AUC in the TIC analysis in the CEUS mode. CDU is more sensitive in detecting feeding vessels than CEUS, while CEUS can better visualize the origins of feeding vessels.

Для цитирования: Gu Guangchao, Zhang Xiaoyan, Shen Junyue, Gulidanna Shayan, Gao Qiong, Shao Jiang, Liu Bao, Zhang Bo, Zheng Yuehong. Comparison of Contrast-Enhanced Ultrasonography to Color Doppler Ultrasound in Evaluation of Carotid Body Tumors.. Frontiers in oncology. 2022. DOI: 10.3389/fonc.2022.872890.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.