DOI 10.1177/08927790251389962

Медицинская библиотека · Journal of endourology · 2025

Comparison of Frictional Forces Among Ureteral Access Sheaths and Urinary Tract Dilators in a Novel Ex Vivo Porcine Ureteral Model.

Camp Brandon, Cumpanas Andrei D, Moreno-Ruiz Kevin, Vu Thao N, Tano Zachary E, Altamirano Jaime, Gorgen Antonio R H, Ali Aymon, Gao Bruce M, Jiang Pengbo, Patel Roshan M, Landman Jaime, Martini Ashlie, Clayman Ralph V

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

Introduction

Surface lubricity is a desirable characteristic of ureteral access sheaths (UAS), allowing for atraumatic navigation of the urinary tract. This study aimed to characterize the surface lubricity of commonly available 14F UAS, as well as 14F ureteral and urethral dilators, using a tribometer and a novel ex vivo ureteral tissue holder. Materials and

Methods

The surface lubricity of 12, 14F urological catheters (8 UAS, 1 urethral dilator, and 3 ureteral dilators) was tested in six ex vivo Yorkshire porcine ureteral segments. A tribometer, capable of measuring force in thousandths of a Newton (N), measured frictional force during insertion and retraction cycles under a 4.0 N normal force. The coefficient of friction (COF) was calculated. Individual and cumulative lubricity scores (ILS and CLS) were calculated based on COF percentiles, enabling a comparative evaluation of device performance. Surface roughness was assessed via white light interferometry.

Results

The Cook Flexor® UAS consistently demonstrated the highest lubricity, as reflected in its CLS for both insertion and retraction phases, followed closely by the Wellead ClearPetra® UAS. Dilators, including the Cook AQ hydrophilic-coated ureteral and urethral dilators, exhibited higher COF values, correlating with greater surface roughness. COF values were significantly higher during retraction cycles than insertion cycles (p < 0.05). Variability in COF across devices underscored the influence of surface properties on lubricity.

Conclusions

Among all the UAS/dilators tested, the Cook Flexor® AQ-coated UAS and Wellead ClearPetra UAS had the best lubricity. Catheter retraction resulted in higher COF than catheter advancement.

Для цитирования: Camp Brandon, Cumpanas Andrei D, Moreno-Ruiz Kevin, Vu Thao N, Tano Zachary E, Altamirano Jaime, Gorgen Antonio R H, Ali Aymon, Gao Bruce M, Jiang Pengbo, Patel Roshan M, Landman Jaime, Martini Ashlie, Clayman Ralph V. Comparison of Frictional Forces Among Ureteral Access Sheaths and Urinary Tract Dilators in a Novel Ex Vivo Porcine Ureteral Model.. Journal of endourology. 2025. DOI: 10.1177/08927790251389962.