Pharmacological Modulation of Intrarenal Pressure in a Porcine Model Using a Novel Isoprenaline-Eluting Guidewire.
Аннотация
Introduction
Several complications of retrograde intrarenal surgery have been attributed to inadvertent increases in intrarenal pressure. We recently described the development of an innovative isoprenaline-eluting guidewire (IsoWire). The objective of this study was to investigate the impact of this IsoWire on the intrarenal pressure and evaluate its safety. Materials and
Methods
This study was performed in 17 renal units using a porcine model. As controls, the intrarenal pressure, heart rate, and mean arterial pressure were measured for a duration of six minutes with a standard guidewire placed in the renal pelvis. For the experiment, the conventional guidewire was substituted with the IsoWire and the same parameters were measured. Blood samples were taken at one-minute intervals to measure plasma isoprenaline levels. This procedure was repeated on the opposite side.
Results
The mean intrarenal pressure reduction was 29% (95% CI: 13%-53%). The mean isoprenaline effect time was 174 seconds. No changes in heart rate (p = .908) or mean arterial pressure (p = .749) were recorded after IsoWire insertion. Plasma isoprenaline levels were below the quantitation threshold. Isoprenaline concentrations in the plasma were below the quantification threshold. Ureteroscopy revealed no ureteral lesions.
Conclusions
The IsoWire demonstrated a safe and effective reduction of intrarenal pressure. Additional research is necessary to determine whether ureteral smooth muscle relaxation generated by isoprenaline facilitates easier insertion of a ureteral access sheath, decreases the incidence of ureteral access sheath related ureteral lesions, or even encourage the practice of sheathless retrograde intrarenal surgery.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.