Does Duty Cycle Modification Affect Temperature During Thulium Fiber Laser Lithotripsy?

Richards Kallan, Hanessian Katya, Albaghli Ali, Martin Gabriel, Jhang Daniel, Battikha Adel, Ghoulian Joshua, Okhunov Zham, Baldwin D Duane

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

OBJECTIVES

Although the thulium fiber laser (TFL, 1940 nm) efficiently fragments stones, concerns have been raised regarding heat production. Recently, a TFL with a modified duty cycle (MDC) was designed to reduce heat generation. The purpose of this study was to compare heat generation between the MDC and conventional TFL.

METHODS

Ten millimeter BegoStone phantoms were placed in the proximal ureter of a kidney and ureter model. The model was submerged in a 35°C saline bath, with continuous saline irrigation (22°C) maintained at 15 mL/min through a ureteroscope. Temperature was measured using a needle thermocouple. Five trials of 60 s continuous laser activation were performed for each power setting: 3 W (0.3 J/10 Hz), 10 W (1 J/10 Hz), 20 W (1 J/20 Hz) and 30 W (0.6 J/50 Hz). Thermal dose was calculated as cumulative equivalent minutes at 43°C (CEM₄₃; injury threshold> 20). Kruskal-Wallis and Mann-Whitney U tests were used for statistical analysis with p < 0.05 being significant.

RESULTS

Both lasers produced maximum temperatures (Tmax) of 27°C at 3 W. The MDC TFL generated a significantly lower Tmax than the conventional TFL at 10 W (36.9°C vs 42.1°C, p < 0.001), 20 W (38.1°C vs 44.6°C, p < 0.001), and 30 W (52.5°C vs 63.6°C, p < 0.05). CEM₄₃ for the MDC and conventional TFL at 20 W were (0.00 vs 0.90, respectively), and at 30 W (425 vs 275,919, respectively).

CONCLUSIONS

The MDC TFL generated lower temperatures at 10, 20, and 30 W. Future studies are necessary to evaluate stone fragmentation efficiency of the MDC TFL.

Для цитирования: Richards Kallan, Hanessian Katya, Albaghli Ali, Martin Gabriel, Jhang Daniel, Battikha Adel, Ghoulian Joshua, Okhunov Zham, Baldwin D Duane. Does Duty Cycle Modification Affect Temperature During Thulium Fiber Laser Lithotripsy?. Lasers in surgery and medicine. 2025. DOI: 10.1002/lsm.70067.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.