Медицинская библиотека · Zhonghua yu fang yi xue za zhi [Chinese journal of preventive medicine] · 2015

[Development and comparative evaluation of up-converting phosphor technology based lateral flow assay for rapid detection of Yersinia pestis, Bacillus anthracis spore and Brucella spp].

Li Chunfeng, Zhang Pingping, Wang Xiaoying, Liu Xiao, Zhao Yong, Sun Chongyun, Wang Chengbin, Yang Ruifu, Zhou Lei

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To develop an up-converting phosphor technology based lateral flow (UPT-LF) assay for rapid and quantitative detection of Yersinia pestis, Bacillus anthracis spore and Brucella spp.and make the comparison with BioThreat Alert (BTA) test strips (Tetracore Inc., USA).

METHODS

Using up-converting phosphor nano-particles (UCP-NPs) as the bio-marker, three double-antibody-sandwich model based UPT-LF strips including Plague-UPT-LF, Anthrax-UPT-LF, Brucella-UPT-LF were prepared and its sensitivity, accuracy, linearity and specificity were determined by detecting 10(10), 10(9), 10(8), 10(7), 10(6), 10(5) and 0 CFU/ml series of concentrations of Y.pestis, B.anthracis, Brucella standards and other 27 kinds of 10(9) CFU/ml series of contrations of bacteria strains.Furthermore, the speed, sensitivity and accuracy of bacteria standards and simulated sample detection were compared between UPT-LF and BTA system.

RESULTS

The detection limit of Plague-UPT-LF, Anthrax-UPT-LF and Brucella-LF was 10(5) CFU/ml. The CV of series of bacteria concentrations was ≤ 15%, and the r between lg (T/C-cut-off) and lg (concentration) was 0.996,0.998 and 0.999 (F values were 1 647.57, 743.51 and 1 822.17. All the P values were <0.001), respectively. The specificity of Plague-UPT-LF and Brucella-LF were excellent, while that of Anthrax-UPT-LF was a little bit regretful because of non-specific reaction with two isolates of B. subtilis and one B.cereus. On-site evaluation showed the detection time of UPT-LF for all Y.pestis, B.anthracis spore and Brucella spp.was 33, 36 and 37 min, while BTA was 115, 115 and 111 min, which revealed the higher detection speed and sensitivity of UPT-LF comparing with BTA. The negative rate of two methods for blank standard was both 5/5, the sensitivity of UPT-LF for Y.pestis,B.anthracis spore and Brucella spp. was all 10(5) CFU/ml, then BTA was 10(6), 10(6) and 10(5) CFU/ml, respectively. The detection rate of UPT-LF for all three bacteria analog positive samples was 16/16, while BTA for B.anthracis was 7/16 only.

CONCLUSION

The good performance including rapidness, simplicity and high sensitivity will bring the bright future of UPT-LF to be broadly used on-site as first response to bio-terrorism.

Для цитирования: Li Chunfeng, Zhang Pingping, Wang Xiaoying, Liu Xiao, Zhao Yong, Sun Chongyun, Wang Chengbin, Yang Ruifu, Zhou Lei. [Development and comparative evaluation of up-converting phosphor technology based lateral flow assay for rapid detection of Yersinia pestis, Bacillus anthracis spore and Brucella spp].. Zhonghua yu fang yi xue za zhi [Chinese journal of preventive medicine]. 2015.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.