The Consistency of Plasma Catecholamines and Metanephrines Analysis in China: A Comparison of Three Mainstream Commercially Available Liquid Chromatography Tandem Mass Spectrometry Kit Methods.

Qin Nijiao, Liu Qingxiang, Liu Zhenni, Li Ziyang, Li Binghui, Guo Zhe, Chen Zhiyang, Wu Weiming, Hu Min, Long Qichen

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

OBJECTIVE

The consistency of plasma catecholamines and metanephrines measurements is crucial for the screening and diagnosis in pheochromocytomas and paragangliomas (PPGLs). There are many commercially available kits for plasma catecholamines and metanephrines testing, but their performance and consistency information are unknown. This study evaluated the performance of three mainstream commercially available LC-MS/MS kit methods and evaluated their consistency.

METHODS

Performance validation of three kit methods was performed according to Clinical Laboratory Standards Institute (CLSI) document C-62A. 40 plasma samples from patients tested for PPGLs were analyzed in three mainstream commercially available kit methods. Agreement of test results from different kit methods was assessed by Passing-Bablok regression and Bland-Altman analysis. Recalibration using uniform calibrations was performed to identify the possible calibration difference.

RESULTS

The performance of three mainstream commercially available LC-MS/MS kit methods was satisfied. The mean deviations ranges were as follows: epinephrine(E), -41.1% ~ -2.4%; norepinephrine (NE), -34.3% ~ -19.1%; metanephrine (MN), -6.5% ~ 0.3% and normetanephrine (NMN), -2.3% to 0.0%. After recalibration, the mean deviation ranges of the E and NE were reduced remarkably (E: -18% to -2.4%, NE: -19.1% to 8.0%).

CONCLUSIONS

For MN and NMN, the consistency between the three commercially available kit methods is acceptable. The results for the E and NE showed unsatisfying comparability for almost all method pairs. The application of uniform calibrators could improve consistency among the mainstream systems for the measurement of plasma catecholamines.

Для цитирования: Qin Nijiao, Liu Qingxiang, Liu Zhenni, Li Ziyang, Li Binghui, Guo Zhe, Chen Zhiyang, Wu Weiming, Hu Min, Long Qichen. The Consistency of Plasma Catecholamines and Metanephrines Analysis in China: A Comparison of Three Mainstream Commercially Available Liquid Chromatography Tandem Mass Spectrometry Kit Methods.. Annals of clinical and laboratory science. 2025.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.