Measurement of exhaled nitric oxide: comparison of different analysers.
Аннотация
BACKGROUND AND
OBJECTIVE
Exhaled nitric oxide (NO) is used as a surrogate marker to monitor eosinophilic airway inflammation, assist in diagnosis, and support treatment decisions for asthma patients. The aim of this study was to compare five NO analysers: Medisoft (M), Aerocrine Niox (N), Aerocrine Niox flex (NF), Aerocrino Niox mino (NM) and EcoMedics (E).
METHODS
In 85 subjects (25 asthma patients, 25 COPD patients, 35 healthy volunteers; median age 36.5 years, range 23-79, 47% female), three NO measurements per individual were performed, using each analyser according to the American Thoracic Society/European Respiratory Society guidelines. Subjects evaluated the devices on the basis of comfort and ease of performing the measurement.
RESULTS
Median NO concentrations were 24.8 parts per billion (ppb) (range 6.3-262.7) for M, 14.5 ppb (0.0-196.6) for N, 15.2 ppb (5.6-67.7) for NF, 16.3 ppb (5.0-208.3) for NM, and 13.1 ppb (4.0-103.0) for E. There was significant correlation among the absolute NO values measured with all the devices (0.743 for M vs NF < r < 0.979 for N vs NF). Bland-Altman plots suggested an acceptable degree of agreement among the results obtained with the three Niox analysers. Measurements obtained with the Medisoft and Ecomedics analysers were not directly comparable with those obtained with the other devices. The greatest differences in absolute NO levels for individual patients were between measurements with the M and E analysers (8.3 ppb, range 27.4-159.7, P < 0.001). Acceptance of the measurements by patients was high, independent of the device used.
CONCLUSIONS
Exhaled NO values obtained with different devices were not directly comparable and may differ to a clinically relevant extent, depending on which device is used.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.