Kreminger Judith E, Told Reinhard, Zafeiri Elena, Reumueller Adrian, Dunavoelgyi Roman, Sacu Stefan
Аннотация
INTRODUCTION: This study aimed to evaluate choroidal nevi (CN) volume as assessed by widefield optical coherence tomography (WF-OCT) and by ultrasound (US)-based mathematical approximations.
METHODS: In a prospective, exploratory study, CN of consecutive patients were imaged with WF-OCT (VG 200D, Intalight Inc., San Jose, California, USA) and US (Quantel Absolu, Cournon, D'Auvergne Cedex, France). Volume was assessed by CN annotation on 64 B-scans on WF-OCT and two approximations based on US measurements (thickness, largest and perpendicular basal diameter (LBD; PBD)): (a) π/6 × thickness × LBD × PBD and (b) 2π/3 × [(LBD + PBD)/4]2 × thickness.
RESULTS: Eighteen nevi of 18 patients were included in the study. Mean volume evaluated on WF-OCT was 15.1 ± 24.1 mm3. US approximations resulted in (a) 23.0 ± 35.0 mm3 and (b) 23.5 ± 35.5 mm3. Intraclass correlation coefficients showed very good agreement between WF-OCT and US measurements (a) of 0.947 (CI 0.860-0.980) and (b) of 0.945 (CI 0.853-0.979). Stepwise backward elimination multivariable linear regression showed an association between difference of volume measurements and lesion thickness and LBD for approximations (a) and (b).
CONCLUSION: WF-OCT and US-based approximations show good interdevice reliability for CN volume in this exploratory study. Thickness and LBD have significant impact on discrepancy of volume measurements with a tendency that small CN are underestimated and large CN are overestimated with US. This potentially limits interchangeable use of US and WF-OCT.
METHODS: In a prospective, exploratory study, CN of consecutive patients were imaged with WF-OCT (VG 200D, Intalight Inc., San Jose, California, USA) and US (Quantel Absolu, Cournon, D'Auvergne Cedex, France). Volume was assessed by CN annotation on 64 B-scans on WF-OCT and two approximations based on US measurements (thickness, largest and perpendicular basal diameter (LBD; PBD)): (a) π/6 × thickness × LBD × PBD and (b) 2π/3 × [(LBD + PBD)/4]2 × thickness.
RESULTS: Eighteen nevi of 18 patients were included in the study. Mean volume evaluated on WF-OCT was 15.1 ± 24.1 mm3. US approximations resulted in (a) 23.0 ± 35.0 mm3 and (b) 23.5 ± 35.5 mm3. Intraclass correlation coefficients showed very good agreement between WF-OCT and US measurements (a) of 0.947 (CI 0.860-0.980) and (b) of 0.945 (CI 0.853-0.979). Stepwise backward elimination multivariable linear regression showed an association between difference of volume measurements and lesion thickness and LBD for approximations (a) and (b).
CONCLUSION: WF-OCT and US-based approximations show good interdevice reliability for CN volume in this exploratory study. Thickness and LBD have significant impact on discrepancy of volume measurements with a tendency that small CN are underestimated and large CN are overestimated with US. This potentially limits interchangeable use of US and WF-OCT.