MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1117/1.jbo.30.6.067003

Медицинская библиотека · Journal of biomedical optics · 2025

Ex vivo investigation of a smart endotracheal tube for identifying esophageal intubation.

Gadsby Brett, Korposh Sergiy, Correia Ricardo, He Chenyang, Hayes-Gill Barrie R, Norris Andrew M, Hardman Jonathan G, Hewson David W, Morgan Stephen P

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

SIGNIFICANCE: Unrecognized intubation of the esophagus instead of the trachea results in rapid and severe consequences for the patient. Utilizing the spectral properties of the tissues could reduce incidents of these events.

AIM: We aim to investigate the design and implementation of a smart endotracheal tube (ETT) with integrated optical fiber sensors to distinguish esophageal and tracheal tissues.

APPROACH: Computational methods are investigated to characterize and classify nine pairs of ex vivo porcine organs using spectral properties. Two classifiers [ K -nearest neighbor and linear discriminant analysis (LDA)] are investigated.

RESULTS

Of the tissues sampled, 100% are correctly distinguished, with LDA being the preferred choice when considering both performance and applicability.

CONCLUSIONS

In clinical practice, this approach offers a method for confirming correct tracheal intubation using the spectral properties of the tissues, performed in a single step with no other invasive medical device than the ETT required to detect the spectral measurements.

Для цитирования: Gadsby Brett, Korposh Sergiy, Correia Ricardo, He Chenyang, Hayes-Gill Barrie R, Norris Andrew M, Hardman Jonathan G, Hewson David W, Morgan Stephen P. Ex vivo investigation of a smart endotracheal tube for identifying esophageal intubation.. Journal of biomedical optics. 2025. DOI: 10.1117/1.jbo.30.6.067003.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.