How to be sure of a complete removal of Essure® implant? A radiological study of Essure® microinserts.

Chene Gautier, Formont Charlotte, Sacdpraseuth Aphynia, Puyol Lea, Miguet-Bensouda Chloe, Moret Stephanie, Cerruto Emanuele, Nohuz Erdogan

СтатьяПеревод в очереди

Аннотация

PURPOSE: Laparoscopic cornuectomy, one of the most performed surgical techniques to remove Essure® implants, is associated with a risk of fragmentation of the device. We aimed to assess the diagnostic accuracy of the intraoperative radiograph of the removed Essure® microinserts compared with intraoperative inspection of the device.

STUDY

DESIGN

In this French retrospective cohort study including 66 patients, two experienced gynecologists and two residents evaluated the intraoperative radiograph in a blind manner. Diagnostic accuracy was calculated for each observer.

RESULTS

The global sensitivity and specificity for intraoperative radiograph to detect a residual fragment for both senior gynecologists and residents were, respectively 87.5 % (±3.3) and 80.0 % (+/12.2). Pooled positive (PPV) and negative predictive values (NPV) for intraoperative radiograph were respectively 99.2 % (±0.5) and 21.6 % (±6.4). There was a statistically significant difference in intraoperative radiograph analysis between senior gynecologists and residents (sensitivity: 91.4 %±0.8 vs 79.6 %±1.5, p = 0.02; specificity: 100% vs 50.0 %, p < 0.0001; PPV: 100 % vs 98.0 % ±0.1, p = 0.0007; NPV: 26.8 % ±1.8 vs 7.2 % ±0.5, p = 0.009). Sensitivity and specificity of the intraoperative inspection of the removal specimen was respectively 95.3 % and 100 % (PPV and NPV were, respectively 100 and 40 %). There were no statistically significant differences in diagnostic accuracy between intraoperative radiographs analysis and intraoperative inspection of the device for the senior gynecologists.

CONCLUSIONS

Intraoperative radiograph of the removed Essure® implants analysis is a difficult skill needing specific training. Postoperative pelvic X-ray is still recommended in clinical practice after laparoscopic cornuectomy.

Для цитирования: Chene Gautier, Formont Charlotte, Sacdpraseuth Aphynia, Puyol Lea, Miguet-Bensouda Chloe, Moret Stephanie, Cerruto Emanuele, Nohuz Erdogan. How to be sure of a complete removal of Essure® implant? A radiological study of Essure® microinserts.. Journal of gynecology obstetrics and human reproduction. 2024. DOI: 10.1016/j.jogoh.2024.102823.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.