Intrarenal Pressure Measured Using a Novel Flexible Ureteroscope with Pressure Sensing Capabilities: A Study of the Effects of Ureteral Access Sheath, Irrigation, and Working Channel Accessories.

Chew Ben H, Shalabi Nabil, Herout Roman, Reicherz Alina, Wong K F Victor, Searles Kyle, Bhojani Naeem

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

Introduction

Intrarenal pressure is emerging as an important monitoring parameter during flexible ureteroscopy (fURS). We measure how intraoperative conditions affect intrarenal pressure using a novel pressure sensing ureteroscope.

Methods

The LithoVue Elite (LVE) single-use digital flexible ureteroscope (Boston Scientific) is the first commercial ureteroscope that senses pressure at its tip. LVE was used in a porcine model to measure intrarenal pressure with and without a ureteral access sheath (UAS) with various sizes and placement locations, irrigation methods, and working channel accessories. LVE pressure accuracy was measured in a bench model. This abstract shows the least-square means from multiway analysis of variances used for analysis.

Results

Intrarenal pressures were the highest without a UAS (64 mm Hg), followed by the 11/13 UAS (51 mm Hg) and the 12/14 and 13/15, which were not statistically different (39-40 mm Hg). The pressures were highest with UASs placed at the ureteropelvic junction (61 mm Hg), and lowest if placed in the renal pelvis (24 mm Hg). Irrigation methods showed the highest pressures with syringe (57 mm Hg), while irrigation bags (pressurized at 150-300 mm Hg) produced 43 to 46 mm Hg and 25 mm Hg when applied with 80 cm of gravity. Placing a 200 μm laser fiber reduced pressures from 44 to 41 mm Hg. Finally, the bench model showed that LVE was 96% accurate up to 300 mm Hg.

Conclusion

Intrarenal pressure significantly varied based on UAS sizes, placement, and irrigation methods. Accordingly, fURS with LVE is poised to be an invaluable tool for clinical decision-making and future studies of intrarenal pressure.

Для цитирования: Chew Ben H, Shalabi Nabil, Herout Roman, Reicherz Alina, Wong K F Victor, Searles Kyle, Bhojani Naeem. Intrarenal Pressure Measured Using a Novel Flexible Ureteroscope with Pressure Sensing Capabilities: A Study of the Effects of Ureteral Access Sheath, Irrigation, and Working Channel Accessories.. Journal of endourology. 2023. DOI: 10.1089/end.2022.0841.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.