Impact of rectal gas evacuation on dosimetry and applicator displacement in cervical cancer brachytherapy.
Аннотация
OBJECTIVE
This study aimed to evaluate the impact of rectal gas evacuation on organ-at-risk (OAR) volumes, dose-volume histogram (DVH) parameters, and applicator displacement during cervical cancer brachytherapy.
METHODS
Twenty-one cervical cancer patients who received three-dimensional brachytherapy at our center between November and December 2024 and presented with rectal gas were retrospectively included. Planning computed tomography (CT) images were acquired before and after rectal gas evacuation to evaluate changes in rectal and bladder volumes, as well as radiation dose variations to OARs (bladder, rectum, sigmoid, and small intestine). Dosimetric parameters analyzed comprised D0.1cc, D1cc, D2cc, and D5cc (minimum doses delivered to the most irradiated 0.1, 1, 2, and 5 cm3 of the OARs, respectively), as well as Dmax (maximum dose) and Dmean (mean dose). Displacements of the applicator tip and cervical stopper were quantified using a coordinate system based on pelvic bony landmarks.
RESULTS
Rectal volume decreased by 40.1% after gas evacuation, while bladder volume increased by 18.2%. D0.1cc, D1cc, D2cc, D5cc, and Dmax in the rectum decreased significantly (P < 0.001) after gas evacuation, whereas no significant changes were observed in the DVH parameters of the other OARs. The mean displacements of the applicator tip and cervical stopper were 5.86 ± 3.64 mm and 4.23 ± 3.30 mm, respectively.
CONCLUSIONS
Rectal gas evacuation results in a statistically significant and clinically relevant reduction in rectal volume and rectal dose, underscoring its importance as a routine clinical procedure. However, as it may induce millimeter-level applicator displacement with clinically measurable dosimetric consequences, careful monitoring is warranted, with post-evacuation replanning or, if necessary, applicator adjustment.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.