Evaluation of Venturi Adapters for CPAP.
Аннотация
Background
CPAP therapy has been shown to effectively reduce the risk of obstructive apnea and oxygen desaturation in procedural sedation. An economical approach to administering CPAP is through a Venturi adapter, which amplifies oxygen flows to produce pressures suitable for CPAP. The objective of this study was to evaluate the performance of four Venturi adapters across clinically relevant metrics including pressure-generation efficiency, inspired oxygen concentration (FIO2), dilution of the end-tidal carbon dioxide (EtCO2) sample for capnometry, and noise levels produced.
Methods
A benchtop simulator was used to evaluate Venturi adapters. The simulator was composed of a test lung, corrugated tubing to simulate a trachea, a silicone-molded face, an air-cushioned mask strapped to the face, a high-pressure oxygen source, and several Venturi adapters. The oxygen flows required to produce specific airway pressures were measured to evaluate pressure-generation efficiency, and the noise levels produced by the Venturi adapters at these pressures were recorded. CO2 was infused into the test lung while the lung was ventilated, and the EtCO2 measured at the mask was compared with a reference measurement taken from the trachea. FIO2 was evaluated by ventilating the test lung and observing the steady-state oxygen concentration in the lung.
Results
Each of the Venturi adapter designs provides their unique strengths and weaknesses. There appears to be a direct trade-off between high pressure-generation efficiency and high FIO2. There also appears to be a direct trade-off between high FIO2 and the extent of dilution in the EtCO2 sample. Some Venturi adapters produce high noise levels, which might result in patient discomfort and clinician unwillingness to use.
Conclusion
Variations in Venturi adapter designs create different clinical benefits. Clinicians should choose the Venturi adapter to best suit their present clinical needs.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.