Comparison of Autofluorescence With Near-Infrared Fluorescence Imaging Between Primary and Secondary Hyperparathyroidism.

Takeuchi Mika, Takahashi Takeshi, Shodo Ryusuke, Ota Hisayuki, Ueki Yushi, Yamazaki Keisuke, Horii Arata

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

To examine the role of autofluorescence (AF) monitoring with near-infrared fluorescence imaging (NIFI) in identifying parathyroid lesions in primary or secondary hyperparathyroidism (P-HPT or S-HPT) surgeries.

STUDY

DESIGN

Prospective study.

METHODS

Twelve lesions each were resected from 12 and 3 patients with P-HPT and S-HPT, respectively. The mean and maximum AF intensities of the lesions normalized to that of the thyroid tissue for in situ and ex vivo preparations were compared between P-HPT and S-HPT. Subjective visual classifications of AF intensity were compared with postoperative quantitative assessments. The unevenness of AF distribution inside the lesions was assessed by determining the ratio of maximum to mean AF intensity and comparing them with the corresponding ratio for normal parathyroid glands (PGs).

RESULTS

In all quantitative comparisons (in situ/ex vivo, mean, and maximum AF), AF intensities of P-HPT were stronger than those of S-PHT. The AF-positive rate in in situ subjective visual classification was higher for P-HPT (100% vs. 33%). Subjective visual classifications showed a positive correlation with AF intensities. The ratio of maximum to mean AF was higher in P-HPT and S-HPT than in normal PGs.

CONCLUSIONS

For P-HPT, AF intensity in both in situ and ex vivo preparations was sufficiently high and correlated with the subjective visual classification, suggesting that NIFI may be useful for confirming P-HPT lesions. In contrast, NIFI may have only a minor role in S-HPT surgeries owing to the weak-AF of S-HPT lesions. HPT lesions show an uneven AF intensity distribution compared with normal PGs.

LEVEL OF EVIDENCE: 3 Laryngoscope, 131:E2097-E2104, 2021.

Для цитирования: Takeuchi Mika, Takahashi Takeshi, Shodo Ryusuke, Ota Hisayuki, Ueki Yushi, Yamazaki Keisuke, Horii Arata. Comparison of Autofluorescence With Near-Infrared Fluorescence Imaging Between Primary and Secondary Hyperparathyroidism.. The Laryngoscope. 2020. DOI: 10.1002/lary.29310.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.