Flexible in-cavity MRI receiving coil for ultra-high-resolution imaging of the pituitary gland.

Liu Siyuan, Patel Kunal S, Peeters Sophie, Lin Jiahao, DiRisio Aislyn C, Vinters Harry V, Candler Rob N, Sung Kyunghyun, Bergsneider Marvin

СтатьяПоддержка NIH: внешняя программаПеревод в очереди

Аннотация

OBJECTIVE

The objective of this study was the preclinical design and construction of a flexible intrasphenoid coil aiming for submillimeter resolution of the human pituitary gland.

METHODS

Sphenoid sinus measurements determined coil design constraints for use in > 95% of adult patients. Temperature safety parameters were tested. The 2-cm-diameter coil prototype was positioned in the sphenoid sinus of cadaveric human heads utilizing the transnasal endoscopic approach that is used clinically. Signal-to-noise ratio (SNR) was estimated for the transnasal coil prototype compared with a standard clinical head coil. One cadaveric pituitary gland was explanted and histologically examined for correlation to the imaging findings.

RESULTS

With the coil positioned directly atop the sella turcica at a 0° angle of the B0 static field, the craniocaudal distance (21.2 ± 0.8 mm) was the limiting constraint. Phantom experiments showed no detectable change in temperature at two sites over 15 minutes. The flexible coil was placed transnasally in cadaveric specimens using an endoscopic approach. The image quality was subjectively superior at higher spatial resolutions relative to that with the commercial 20-channel head coil. An average 17-fold increase in the SNR was achieved within the pituitary gland. Subtle findings visualized only with the transnasal coil had potential pathological correlation with immunohistochemical findings.

CONCLUSIONS

A transnasal radiofrequency coil feasibly provides a 17-fold boost in the SNR at 3 T. The ability to safely improve the quality of pituitary imaging may be helpful in the identification and subsequent resection of small functional pituitary lesions.

Для цитирования: Liu Siyuan, Patel Kunal S, Peeters Sophie, Lin Jiahao, DiRisio Aislyn C, Vinters Harry V, Candler Rob N, Sung Kyunghyun, Bergsneider Marvin. Flexible in-cavity MRI receiving coil for ultra-high-resolution imaging of the pituitary gland.. Journal of neurosurgery. 2023. DOI: 10.3171/2023.7.jns231096.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.