MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.apradiso.2026.112554

Медицинская библиотека · Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026

Quantitative evaluation of patient-emitted radiation in dual-injection Tc-99m MIBI myocardial perfusion scintigraphy: A multiregional, multi-distance, and scenario-based safety analysis.

Şenışık Ahmet Murat, Kökkülünk Handan Tanyıldızı, Yüksel Mahmut

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Patients undergoing Tc-99m MIBI myocardial perfusion scintigraphy (MPS) temporarily act as external photon sources, creating short-term radiation exposure to staff and accompanying persons. Quantitative real-patient data describing multiregional and multi-distance dose-rate behavior under dual-injection protocols remain limited.

PURPOSE: To characterize patient-emitted radiation following dual-injection Tc-99m MIBI MPS and to estimate cumulative doses for realistic occupational and public-contact scenarios.

METHODS

Ambient dose equivalent rates (μSv/h) were measured at four anatomical levels (head, thorax, abdomen, pelvis) and five distances (25-200 cm) at three post-stress time points. Distance dependence was evaluated using power-law regression, and temporal clearance at 100 cm was modeled mono-exponentially to derive region-specific effective half-lives. Scenario-based cumulative doses were calculated using both median (typical) and 95th percentile (conservative) dose rates.

RESULTS

Dose rates decreased consistently with distance in all regions, with excellent regression fits (R2 ≥ 0.995) and empirical attenuation exponents of 0.53-0.76, consistent with extended-source behavior. Effective half-lives ranged from 4.65 to 5.69 h. Scenario-based cumulative exposures remained in the microsievert range, including ≤1.14 μSv for typical technologist contact and ≤9.69 μSv for conservative prolonged caregiver proximity.

CONCLUSION

Dual-injection Tc-99m MIBI MPS results in very low external radiation exposure to staff and accompanying persons. Measured doses are far below regulatory constraints, supporting the adequacy of routine ALARA-based clinical practice without additional restrictions.

Для цитирования: Şenışık Ahmet Murat, Kökkülünk Handan Tanyıldızı, Yüksel Mahmut. Quantitative evaluation of patient-emitted radiation in dual-injection Tc-99m MIBI myocardial perfusion scintigraphy: A multiregional, multi-distance, and scenario-based safety analysis.. Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine. 2026. DOI: 10.1016/j.apradiso.2026.112554.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.