Motion-resolved parametric imaging derived from short dynamic [18F]FDG PET/CT scans.

Artesani Alessia, van Sluis Joyce, Providência Laura, van Snick Johannes H, Slart Riemer H J A, Noordzij Walter, Tsoumpas Charalampos

СтатьяПеревод в очереди

Аннотация

PURPOSE: This study aims to assess the added value of utilizing short-dynamic whole-body PET/CT scans and implementing motion correction before quantifying metabolic rate, offering more insights into physiological processes. While this approach may not be commonly adopted, addressing motion effects is crucial due to their demonstrated potential to cause significant errors in parametric imaging.

METHODS

A 15-minute dynamic FDG PET acquisition protocol was utilized for four lymphoma patients undergoing therapy evaluation. Parametric imaging was obtained using a population-based input function (PBIF) derived from twelve patients with full 65-minute dynamic FDG PET acquisition. AI-based registration methods were employed to correct misalignments between both PET and ACCT and PET-to-PET. Tumour characteristics were assessed using both parametric images and standardized uptake values (SUV).

RESULTS

The motion correction process significantly reduced mismatches between images without significantly altering voxel intensity values, except for SUVmax. Following the alignment of the attenuation correction map with the PET frame, an increase in SUVmax in FDG-avid lymph nodes was observed, indicating its susceptibility to spatial misalignments. In contrast, Patlak Ki parameter was highly sensitive to misalignment across PET frames, that notably altered the Patlak slope. Upon completion of the motion correction process, the parametric representation revealed heterogeneous behaviour among lymph nodes compared to SUV images. Notably, reduced volume of elevated metabolic rate was determined in the mediastinal lymph nodes in contrast with an SUV of 5 g/ml, indicating potential perfusion or inflammation.

CONCLUSIONS

Motion resolved short-dynamic PET can enhance the utility and reliability of parametric imaging, an aspect often overlooked in commercial software.

Для цитирования: Artesani Alessia, van Sluis Joyce, Providência Laura, van Snick Johannes H, Slart Riemer H J A, Noordzij Walter, Tsoumpas Charalampos. Motion-resolved parametric imaging derived from short dynamic [18F]FDG PET/CT scans.. Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB). 2025. DOI: 10.1016/j.ejmp.2025.105010.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.