MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/bco2.473

Медицинская библиотека · BJUI compass · 2024

In silico tools for mechanical analysis of extra- and intra-luminal artificial urinary sphincters.

Mazzucco Gianluca, Pirini Paola, Fontanella Chiara Giulia, Berardo Alice, Mascolini Maria Vittoria, Toniolo Ilaria, Marziale Leonardo, Mazzocchi Tommaso, Lucarini Gioia, Spiezia Nicolò, Carniel Emanuele Luigi

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

To analyse and compare the functionality of extraluminal and intraluminal artificial urinary sphincters (AUSs), an in silico procedure has been defined and applied. Design and reliability assessments of the AUS are typically performed using a clinical approach, which does not provide data on mechanical stimulation of urethral tissues. Mechanical stimulation may determine tissue degeneration, such as urethral atrophy or erosion, the main causes of AUS failure. In silico techniques can provide a quantitative description of stress and strain fields due to the interaction between tissues and AUS and allow investigating an extremely large number of situations, considering different configurations of AUS and urethra.

MATERIALS AND

METHODS

Computational investigations were carried out to evaluate the mechanical reliability of the main extraluminal and intraluminal AUS, AMS 800 and Relief. The lower urinary tract was modelled based on previous experiments. The AUS models took into account the main components that interact with biological tissues. Urethra and AUS models were coupled and used to investigate mechanical stimulation of urethral tissues.

RESULTS

In silico simulations provide quantitative information about the mechanical stimulation of urethral tissue, such as compressive strain and stress and hydrostatic pressure, due to interaction with the AUS. Such mechanical quantities allow a comparison of reliability between extraluminal and intraluminal devices.

CONCLUSIONS

The activities define and demonstrate the effectiveness of a novel in silico approach to the design and reliability assessment of AUS devices, increasing the investigative possibilities and reducing the time, ethical and economic costs.

Для цитирования: Mazzucco Gianluca, Pirini Paola, Fontanella Chiara Giulia, Berardo Alice, Mascolini Maria Vittoria, Toniolo Ilaria, Marziale Leonardo, Mazzocchi Tommaso, Lucarini Gioia, Spiezia Nicolò, Carniel Emanuele Luigi. In silico tools for mechanical analysis of extra- and intra-luminal artificial urinary sphincters.. BJUI compass. 2024. DOI: 10.1002/bco2.473.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.