A comparative in vitro study of the accuracy and precision of different intraoral scanners for implant-supported hybrid prostheses impressions.

İnal Suhan, Uğur Mehmet

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

BACKGROUND

The accuracy and precision of intraoral scanners (IOSs) are critical factors for achieving a passive fit in implant-supported hybrid prostheses (ISHP). However, limited evidence is available regarding the comparative performance of different IOSs in complete-arch implant cases. This in vitro study aimed to simultaneously evaluate and compare the trueness and repeatability of six contemporary IOS systems in capturing mandibular ISHP impressions.

METHODS

A mandibular acrylic model with five implant analogs was scanned using an industrial scanner (Zeiss Comet 6) to obtain a reference dataset. Six IOSs (Primescan, Trios 4, Trios 3, Medit i500, Virtuo Vivo, and Emerald S) were each used to scan the model five times by a single operator under controlled laboratory conditions. All test scans were superimposed on the reference using Geomagic Control X software, and root mean square (RMS) values were calculated to assess accuracy. Precision was determined using the coefficient of variation (CV). Data were analyzed with one-way ANOVA and Games–Howell post-hoc tests (α = 0.05).

RESULTS

Significant differences were observed among scanners (p < 0.05). Primescan exhibited the highest trueness (0.0411 ± 0.0058 mm), followed by Trios 4 and Trios 3, whereas Emerald S demonstrated the lowest trueness (0.056 ± 0.0068 mm). The Trios 3 demonstrated the greatest repeatability (CV = 0.021), while Virtuo Vivo exhibited the lowest (CV = 0.179).

CONCLUSION

Scanner selection significantly influences digital impression accuracy for ISHP. The Primescan and Trios systems provided superior trueness and precision, respectively, highlighting their clinical advantage in full-arch implant rehabilitations. Clinicians should consider the optical technology and demonstrated performance of IOSs when integrating digital workflows for complex implant-supported restorations.

Для цитирования: İnal Suhan, Uğur Mehmet. A comparative in vitro study of the accuracy and precision of different intraoral scanners for implant-supported hybrid prostheses impressions.. BMC oral health. 2026. DOI: 10.1186/s12903-026-07692-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.