DOI 10.1097/j.jcrs.0000000000001860

Медицинская библиотека · Journal of cataract and refractive surgery · 2026

Halo assessment in intraocular lenses through high dynamic range images.

García-Pedreño Clara, Tabernero Juan, Ginis Harilaos, Hervella Lucía, Artal Pablo

СтатьяПеревод в очереди

Аннотация

PURPOSE: To analyze the halo formation of several intraocular lenses (IOLs) in the optical bench.

SETTING

University of Murcia, Murcia, Spain.

DESIGN

In vitro study.

METHODS

Light from a green LED passed through a pinhole and was collimated. Each IOL was placed within a realistic model eye having a PMMA cornea with a physiological amount of spherical aberration and a 4.5 mm aperture. A digital camera sensor acted as the retina and a focus tunable lens was used to change the object's vergence (range ±4 diopters). Series of images were captured with different exposure times and fused to get a high dynamic range image. Performance was assessed by analyzing the corresponding halo brightness and size. The tested lenses, that included biconvex and inverted meniscus IOLs, were monofocals, extended depth-of-focus (EDOF), and diffractive trifocals.

RESULTS

Monofocal lenses produced halos with a radius close to 0.4 degrees. The halo radii of the nondiffractive EDOF lenses ranged between 0.45 and 0.63 degrees, whereas diffractive lenses had radii ranging from 0.84 to 1.22 degrees. The halo was generally dimmer for the refractive lenses and brighter for the diffractive. The through-focus images show that the halo size was larger at any defocus position for the diffractive lenses than for the rest of the tested IOLs.

CONCLUSIONS

The diffractive IOLs exhibited a characteristic halo structure. Performance of the inverted meniscus and other nondiffractive lenses (Vivity and Eyhance) was comparable with a monofocal lens. This on-bench test can serve as an indication of the potential impact of photic phenomena on patient satisfaction.

Для цитирования: García-Pedreño Clara, Tabernero Juan, Ginis Harilaos, Hervella Lucía, Artal Pablo. Halo assessment in intraocular lenses through high dynamic range images.. Journal of cataract and refractive surgery. 2026. DOI: 10.1097/j.jcrs.0000000000001860.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.