MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/mao.0000000000004379

Медицинская библиотека · Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2025

Intraoperative Electrical Stapedius Reflex Testing to Monitor Cochlear Nerve Integrity during Simultaneous Vestibular Schwannoma Resection and Cochlear Implantation.

Carlson Matthew L, Daher Ghazal S, Khandalavala Karl R, Lohse Christine M, Neff Brian A, Driscoll Colin L W, Peris-Celda Maria, Van Gompel Jamie J, Link Michael J, Saoji Aniket A

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To compare the utility of intraoperative electrically evoked auditory brainstem response (eABR) and electrically evoked stapedial reflex testing (eSRT) for cochlear nerve integrity monitoring during simultaneous translabyrinthine resection of vestibular schwannoma (VS) and cochlear implantation.

STUDY

DESIGN

Historical cohort study.

SETTING

Tertiary academic referral center.

PATIENTS: Eleven patients (12 cases) who underwent translabyrinthine tumor resection with simultaneous placement of a standard, unmodified cochlear implant (CI).

INTERVENTION: A CI placed at the beginning of the operation before tumor resection to facilitate intraoperative cochlear nerve monitoring through a CI-delivered electrical stimulus.

MAIN OUTCOME MEASURE: Correlation of intraoperative monitoring outcomes with postoperative CI performance.

RESULTS

Among the 12 cases (median age, 47 years), eABR waveforms were present at the conclusion of tumor resection in 10, including 6 that achieved open set speech perception. Of the 2 cases with negative eABR, 1 achieved open set speech perception. By contrast, of the 8 cases with a positive eSRT response, 7 achieved open set speech perception. Of the 4 cases with a negative eSRT response, none achieved open set speech perception. The 3 cases that did not receive auditory stimulation included 2 with positive and 1 with negative eABR; all 3 had a negative eSRT response.

CONCLUSIONS

Intraoperative eSRT using an unmodified CI is a promising new method for monitoring the integrity of the cochlear nerve during simultaneous VS resection. eSRT response at the conclusion of tumor resection strongly correlated with CI open set speech capacity. Although still useful, eABR was less reliable at predicting postoperative hearing outcomes in this series.

Для цитирования: Carlson Matthew L, Daher Ghazal S, Khandalavala Karl R, Lohse Christine M, Neff Brian A, Driscoll Colin L W, Peris-Celda Maria, Van Gompel Jamie J, Link Michael J, Saoji Aniket A. Intraoperative Electrical Stapedius Reflex Testing to Monitor Cochlear Nerve Integrity during Simultaneous Vestibular Schwannoma Resection and Cochlear Implantation.. Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology. 2025. DOI: 10.1097/mao.0000000000004379.