Development of a bespoke phantom to optimize molecular PET imaging of pituitary tumors.

Gillett Daniel, Marsden Daniel, Crawford Rosy, Ballout Safia, MacFarlane James, van der Meulen Merel, Gillett Bethany, Bird Nick, Heard Sarah, Powlson Andrew S, Santarius Thomas, Mannion Richard, Kolias Angelos, Harper Ines, Mendichovszky Iosif A, Aloj Luigi, Cheow Heok, Bashari Waiel, Koulouri Olympia, Gurnell Mark

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Image optimization is a key step in clinical nuclear medicine, and phantoms play an essential role in this process. However, most phantoms do not accurately reflect the complexity of human anatomy, and this presents a particular challenge when imaging endocrine glands to detect small (often subcentimeter) tumors. To address this, we developed a novel phantom for optimization of positron emission tomography (PET) imaging of the human pituitary gland. Using radioactive 3D printing, phantoms were created which mimicked the distribution of 11C-methionine in normal pituitary tissue and in a small tumor embedded in the gland (i.e., with no inactive boundary, thereby reproducing the in vivo situation). In addition, an anatomical phantom, replicating key surrounding structures [based on computed tomography (CT) images from an actual patient], was created using material extrusion 3D printing with specialized filaments that approximated the attenuation properties of bone and soft tissue.

RESULTS

The phantom enabled us to replicate pituitary glands harboring tumors of varying sizes (2, 4 and 6 mm diameters) and differing radioactive concentrations (2 ×, 5 × and 8 × the normal gland). The anatomical phantom successfully approximated the attenuation properties of surrounding bone and soft tissue. Two iterative reconstruction algorithms [ordered subset expectation maximization (OSEM); Bayesian penalized likelihood (BPL)] with a range of reconstruction parameters (e.g., 3, 5, 7 and 9 OSEM iterations with 24 subsets; BPL regularization parameter (β) from 50 to 1000) were tested. Images were analyzed quantitatively and qualitatively by eight expert readers. Quantitatively, signal was the highest using BPL with β = 50; noise was the lowest using BPL with β = 1000; contrast was the highest using BPL with β = 100. The qualitative review found that accuracy and confidence were the highest when using BPL with β = 400.

CONCLUSIONS

The development of a bespoke phantom has allowed the identification of optimal parameters for molecular pituitary imaging: BPL reconstruction with TOF, PSF correction and a β value of 400; in addition, for small (< 4 mm) tumors with low contrast (2:1 or 5:1), sensitivity may be improved using a β value of 100. Together, these findings should increase tumor detection and confidence in reporting scans.

Для цитирования: Gillett Daniel, Marsden Daniel, Crawford Rosy, Ballout Safia, MacFarlane James, van der Meulen Merel, Gillett Bethany, Bird Nick, Heard Sarah, Powlson Andrew S, Santarius Thomas, Mannion Richard, Kolias Angelos, Harper Ines, Mendichovszky Iosif A, Aloj Luigi, Cheow Heok, Bashari Waiel, Koulouri Olympia, Gurnell Mark. Development of a bespoke phantom to optimize molecular PET imaging of pituitary tumors.. EJNMMI physics. 2023. DOI: 10.1186/s40658-023-00552-9.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.