MedElevator Войти в исследования
← Поиск публикаций

DOI 10.5152/tud.2025.24100

Медицинская библиотека · Urology research & practice · 2025

Comparison of Temperature Changes Between Holmium-YAG Laser and Thulium Fiber Laser in an In Vitro Setting.

Patel Vaibhavkumar, Raghuvanshi Kshitij, Jadeja Krushnadevsinh, Chaudhari Rajeev

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

We aimed to evaluate and compare the rise in the temperature for the safety of the ureter and kidney parenchyma when firing the holmium laser and the thulium fiber laser (TFL).

METHODS

We performed a laboratory experiment to measure the rise in temperature upon firing holmium laser and a TFL in a 10 cm3 transparent test tube in an outdoor environment and then in a container with normal saline.

RESULTS

In a 10 cc test tube with static water at 25°C, the rise in temperature with holmium and TFL depends on the firing time, keeping power constant. On continuous firing for 10, 20, and 30 seconds, the temperature rose to 35.1, 42.7, and 53.2°C with holmium. The temperature went up to 38.3, 46.8, and 55.4°C when TFL was used. The power was kept at 10 W for both types of lasers. The temperature rise was relatively low when the test tube was immersed in a water container. It was even lower with irrigation.

CONCLUSION

The rise in the temperature reaches to hazardous level in static water with a firing time of 30 seconds, which is almost equivalent to holmium and TFL. Thus, while using lasers in RIRS and ureteroscopy, the firing time should not exceed 20 seconds and adequate irrigation is required to prevent damage to the tissues. Also, the rise in temperature was almost equivalent to holmium and TFL; hence, TFL can be safely used in laser lithotripsy in any part of the genitourinary (GU) system.

Для цитирования: Patel Vaibhavkumar, Raghuvanshi Kshitij, Jadeja Krushnadevsinh, Chaudhari Rajeev. Comparison of Temperature Changes Between Holmium-YAG Laser and Thulium Fiber Laser in an In Vitro Setting.. Urology research & practice. 2025. DOI: 10.5152/tud.2025.24100.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.