Beam profile of the flashlamp pumped pulsed dye laser: support for overlap of exposure spots.
Аннотация
The degree to which individual practitioners overlap exposure spots of the flashlamp pumped pulsed dye laser when treating large vascular lesions appears to be a matter of personal preference. There are few data to support one approach over another. We have measured the beam profile of this laser at two different energy fluences using a photodiode and a digitizing storage oscilloscope. Results show that for each energy fluence, the beam intensity decreases from the center to the outside rim of the beam. A plot of beam intensity versus the radius of the treatment field approaches a truncated normal distribution with a full width at half-maximum of 4.7-4.8 mm. These results help explain why clinical overlap of exposure spots can be safely undertaken and provide a rationale for recommendations concerning the degree of overlap needed to expedite treatment while still minimizing adverse reactions.