MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3390/jcm12216740

Медицинская библиотека · Journal of clinical medicine · 2023

Bleb Morphology on Anterior-Segment Optical Coherence Tomography after XEN Gel Stent Implantation.

Wy Seoyoung, Shin Young In, Kim Young Kook, Jeoung Jin Wook, Park Ki Ho

СтатьяПеревод в очереди

Аннотация

We investigated the internal morphology of filtration blebs after XEN gel stent implantation using anterior segment optical coherence tomography (AS-OCT) and identified factors related to intraocular pressure (IOP) postoperatively. Eighteen eyes of 18 patients who had undergone XEN gel stent implantation were analyzed. Blebs were imaged using Visante OCT (Carl Zeiss Meditec AG, Germany) at 6 months and 1 year after surgery and evaluated for quantitative parameters including bleb height, maximum height of internal cavity, maximum bleb wall thickness, and maximum bleb epithelial thickness. Subjects were classified into two groups according to the presence or absence of a definite internal cavity between the conjunctiva and sclera using AS-OCT imaging. Nine eyes (50%) were assigned to the internal cavity group and 9 (50%) to the uniform group. Postoperative IOP was significantly lower in the internal cavity group than in the uniform group both at 6 months and 1 year after surgery (p = 0.024 and p = 0.040). Postoperative IOP showed statistically significant negative correlations with bleb height and the height of the internal cavity (Spearman correlation coefficient r = -0.518, p = 0.028 and r = -0.453, p = 0.034, respectively). AS-OCT facilitates analysis of bleb morphology after XEN gel stent implantation. A larger height of the internal cavity of the bleb appeared to correlate with lower IOP after XEN implantation.

Для цитирования: Wy Seoyoung, Shin Young In, Kim Young Kook, Jeoung Jin Wook, Park Ki Ho. Bleb Morphology on Anterior-Segment Optical Coherence Tomography after XEN Gel Stent Implantation.. Journal of clinical medicine. 2023. DOI: 10.3390/jcm12216740.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.