Pre-clinical and clinical evaluation of a surface plasmon field-enhanced fluorescence spectroscopy (SPFS) antigen test for detecting SARS-CoV-2.

Ashizawa Nobuyuki, Takazono Takahiro, Ota Kenji, Ito Yuya, Hirayama Tatsuro, Takeda Kazuaki, Ide Shotaro, Iwanaga Naoki, Fujita Ayumi, Tashiro Masato, Hosogaya Naoki, Akamatsu Norihiko, Kosai Kosuke, Tanaka Takeshi, Kobayashi Hironori, Yamauchi Ryosuke, Segawa Chie, Koizumi Hikaru, Taka Natsumi, Hishigaki Haretsugu, Yamamoto Kazuko, Izumikawa Koichi, Yanagihara Katsunori, Mukae Hiroshi

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

The diagnostic tools of nucleic acid amplification tests and antigen tests have been extensively employed for the detection of Coronavirus disease 2019 (COVID-19). Although the reverse-transcriptase polymerase chain reaction (RT)-PCR test has high sensitivity and specificity, it is a time-consuming and labor-intensive process. On the other hand, antigen tests are simple and prompt, however, their low sensitivity and potential for false positives have been identified as limitations. In light of these factors, the development of novel tests that combine speed and clinical dependability is a promising prospect.

METHODS

Surface plasmon field-enhanced fluorescence spectroscopy (SPFS) excites chromophores by means of an enhanced electromagnetic field induced on a gold film surface. It enables the highly sensitive measurement of biomarkers in a short and simple 20-min window. In this study, a severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) SPFS-based antigen test targeting the SARS-CoV-2 nucleocapsid protein was performed and evaluated in 25 patients with COVID-19 and 10 non-infected controls.

RESULTS

A positive correlation was observed between antigen levels determined by SPFS and RNA levels determined via RT-PCR. The sensitivity values were 100 %, 92 %, and 62.5 %; and the specificity values were 100 %, 90 %, and 100 %; for nasopharyngeal swabs, nasal swabs, and saliva specimens when the cutoff values were set to 65.1, 0.2, and 1.5 pg/mL, respectively. No clinically problematic cross-reactivity with analogous coronaviruses was observed.

CONCLUSIONS

The SARS-CoV-2 SPFS antigen test showed excellent clinical diagnostic accuracy for nasopharyngeal and nasal swabs, with a rapid turnaround.

Для цитирования: Ashizawa Nobuyuki, Takazono Takahiro, Ota Kenji, Ito Yuya, Hirayama Tatsuro, Takeda Kazuaki, Ide Shotaro, Iwanaga Naoki, Fujita Ayumi, Tashiro Masato, Hosogaya Naoki, Akamatsu Norihiko, Kosai Kosuke, Tanaka Takeshi, Kobayashi Hironori, Yamauchi Ryosuke, Segawa Chie, Koizumi Hikaru, Taka Natsumi, Hishigaki Haretsugu, Yamamoto Kazuko, Izumikawa Koichi, Yanagihara Katsunori, Mukae Hiroshi. Pre-clinical and clinical evaluation of a surface plasmon field-enhanced fluorescence spectroscopy (SPFS) antigen test for detecting SARS-CoV-2.. Journal of infection and chemotherapy : official journal of the Japan Society of Chemotherapy. 2024. DOI: 10.1016/j.jiac.2024.08.019.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.