MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/jgh.70716

Медицинская библиотека · Journal of gastroenterology and hepatology · 2026

A Novel Anchoring Catheter Improves the Success, Efficiency, and Satisfaction of Paracentesis in Refractory Ascites: A Multicenter Randomized Crossover Trial.

Jeon So Hyun, Heo Nae-Yun, Maeng Seon Ah, Yoo Hae Won, Yoo Jeong-Ju, Kim Young Seok, Park Seung Ha, Seo Yeon Seok, Kim Sang Gyune

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Large-volume paracentesis is the standard treatment for refractory ascites, but conventional angiocatheter (CA) use may be limited by kinking and dislodgement. KARAHOC (KH) is a multifenestrated over-the-needle catheter with an anchoring disc designed to improve catheter stability. We compared the efficacy and safety of KH with CA in cirrhotic patients with refractory ascites.

METHODS

This prospective, multicenter, 2 × 2 randomized crossover trial enrolled 87 cirrhotic patients requiring repeated paracentesis at three Korean academic hospitals. Participants underwent paracentesis with both KH and CA in randomized order at least 7 days apart. Successful paracentesis was defined as drainage of ≥ 3 L of ascites. Patient and operator satisfaction were assessed using a visual analog scale.

RESULTS

In the intention-to-treat analysis, success was higher with KH than with CA (87.5% vs. 71.6%; OR 2.78, 95% CI 1.22-6.30; p = 0.018), with shorter median procedure time (71 vs. 80 min; p = 0.018). Per-protocol analysis showed similar results (91.9% vs. 77.0%; p = 0.015). Re-puncture was less frequent with KH but was not statistically significant. Patient and operator satisfaction were significantly higher with KH (both p < 0.001). Adverse events were comparable, with no major complications.

CONCLUSION

KH improved paracentesis success, shortened procedure time, and increased satisfaction without compromising safety, supporting its use as a practical alternative to CA for repeated paracentesis in refractory ascites.

TRIAL REGISTRATION: ClinicalTrials.gov: NCT05578573.

Для цитирования: Jeon So Hyun, Heo Nae-Yun, Maeng Seon Ah, Yoo Hae Won, Yoo Jeong-Ju, Kim Young Seok, Park Seung Ha, Seo Yeon Seok, Kim Sang Gyune. A Novel Anchoring Catheter Improves the Success, Efficiency, and Satisfaction of Paracentesis in Refractory Ascites: A Multicenter Randomized Crossover Trial.. Journal of gastroenterology and hepatology. 2026. DOI: 10.1111/jgh.70716.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.