Biomechanical analysis of a disc prosthesis distal to a scoliosis model.

Quirno Martin, Kamerlink Jonathan R, Valdevit Antonio, Kang Matthew, Yaszay Burt, Duncan Naphysah, Boachie-Adjei Oheneba, Lonner Baron S, Errico Thomas J

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

STUDY

DESIGN

Biomechanical study of bovine spines.

OBJECTIVE

The purpose of this study was to perform a biomechanical test to analyze intervertebral deflections following placement of both 1 and 2 semiconstrained TDRs in the subjacent segments of a long fusion.

SUMMARY OF BACKGROUND DATA: Long-term sequela of long lumbar fusion for scoliosis include adjacent segment disease and flatback syndrome. Total disc replacement (TDR) is a viable option for the treatment of these conditions. Little data has been published regarding the placement of a TDR distal to a scoliosis fusion.

METHODS

Six thoracolumbar bovine spines (T12-S1) were instrumented from T12 to L5, with bilateral pedicle screw fixation at each level. L5-L6 and L6-S1 served as the test levels. One TDR (FlexiCore, Stryker Spine, Allendale, NJ) was initially performed adjacent to the fusion, followed by a subsequent TDR insertion at the last spinal segment. The applied load, total specimen deflection, and local transducer deflections were recorded before and after a TDR at both levels. The results were expressed as a percentage of the intact specimen. Flexion, extension, lateral bending, and torsional deflections were recorded.

RESULTS

There were no significant differences (P > 0.05) in sensor deflection observed at the L5-L6 and L6-S1 levels in the anterior and lateral transducers when compared to intact spines specimens. A similar effect was observed at the L5-L6 and L6-S1 levels in the anterior and lateral transducers when compared to intact or prior L5-L6 and intact L6-S1 constructs.

CONCLUSION

This study has shown that using the FlexiCore system at 1 and/or 2 intervertebral disc spaces caudal to a scoliosis fusion model did not significantly change the sensor deflection at the 2 segments adjacent to a scoliosis fusion construct. Future research will continue to define the clinical setting and patients best suited for management by TDR systems.

Для цитирования: Quirno Martin, Kamerlink Jonathan R, Valdevit Antonio, Kang Matthew, Yaszay Burt, Duncan Naphysah, Boachie-Adjei Oheneba, Lonner Baron S, Errico Thomas J. Biomechanical analysis of a disc prosthesis distal to a scoliosis model.. Spine. 2009. DOI: 10.1097/brs.0b013e3181a8e418.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.