Can a Wireless Full-HD Head Mounted Display System Improve Knee Arthroscopy Performance? - A Randomized Study Using a Knee Simulator.
Рандомизированное исследованиеСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Fang Christian, Mo Pinky, Chan Holy, Cheung Jake, Wong Janus Siu Him, Wong Tak-Man, Mak Yan-Kit, Ching Kathine, Ho Grace, Leung Frankie
Аннотация
INTRODUCTION: Our prototype wireless full-HD Augmented Reality Head-Mounted Display (AR-HMD) aims to eliminate surgeon head turning and reduce theater clutter. Learning and performance versus TV Monitors (TVM) is evaluated in simulated knee arthroscopy.
METHODS: 19 surgeons and 19 novices were randomized into either the control group (A) or intervention group (B) and tasked to perform 5 simulated loose-body retrieval procedures on a bench-top knee arthroscopy simulator. A cross-over study design was adopted whereby subjects alternated between devices during trials 1-3, deemed the "Unfamiliar" phase, and then used the same device consecutively in trials 4-5, to assess performance in a more "Familiarized" state. Measured outcomes were time-to-completion and incidence of bead drops.
RESULTS: In the unfamiliar phase, HMD had 67% longer mean time-to-completion than TVM (194.7 ± 152.6s vs 116.7 ± 78.7s, P < .001). Once familiarized, HMD remained inferior to TVM, with 48% longer completion times (133.8 ± 123.3s vs 90.6 ± 55s, P = .052). Cox regression revealed device type (OR = 0.526, CI 0.391-0.709, P < .001) and number of procedure repetitions (OR = 1.186, CI 1.072-1.311, P = .001) are significantly and independently related to faster time-to-completion. However, experience is not a significant factor (OR = 1.301, CI 0.971-1.741, P = .078). Bead drops were similar between the groups in both unfamiliar (HMD: 27 vs TVM: 22, P = .65) and familiarized phases (HMD: 11 vs TVM: 17, P = .97).
CONCLUSION: Arthroscopic procedures continue to be better performed under conventional TVM. However, similar quality levels can be reached by HMD when given more time. Given the theoretical advantages, further research into improving HMD designs is advocated.
METHODS: 19 surgeons and 19 novices were randomized into either the control group (A) or intervention group (B) and tasked to perform 5 simulated loose-body retrieval procedures on a bench-top knee arthroscopy simulator. A cross-over study design was adopted whereby subjects alternated between devices during trials 1-3, deemed the "Unfamiliar" phase, and then used the same device consecutively in trials 4-5, to assess performance in a more "Familiarized" state. Measured outcomes were time-to-completion and incidence of bead drops.
RESULTS: In the unfamiliar phase, HMD had 67% longer mean time-to-completion than TVM (194.7 ± 152.6s vs 116.7 ± 78.7s, P < .001). Once familiarized, HMD remained inferior to TVM, with 48% longer completion times (133.8 ± 123.3s vs 90.6 ± 55s, P = .052). Cox regression revealed device type (OR = 0.526, CI 0.391-0.709, P < .001) and number of procedure repetitions (OR = 1.186, CI 1.072-1.311, P = .001) are significantly and independently related to faster time-to-completion. However, experience is not a significant factor (OR = 1.301, CI 0.971-1.741, P = .078). Bead drops were similar between the groups in both unfamiliar (HMD: 27 vs TVM: 22, P = .65) and familiarized phases (HMD: 11 vs TVM: 17, P = .97).
CONCLUSION: Arthroscopic procedures continue to be better performed under conventional TVM. However, similar quality levels can be reached by HMD when given more time. Given the theoretical advantages, further research into improving HMD designs is advocated.