Tao Qiuying, Wang Kaiyu, Wen Baohong, Kang Yimeng, Dang Jinghan, Sun Jieping, Niu Xiaoyu, Zhang Mengzhe, Liu Zijun, Wang Weijian, Zhang Yong, Cheng Jingliang

Аннотация

OBJECTIVES: To investigate potential of enhancing image quality, maintaining interobserver consensus, and elevating disease diagnostic efficacy through the implementation of deep learning-based reconstruction (DLR) processing in 3.0 T cervical spine fast magnetic resonance imaging (MRI) images, compared with conventional images.

METHODS: The 3.0 T cervical spine MRI images of 71 volunteers were categorized into two groups: sagittal T2-weighted short T1 inversion recovery without DLR (Sag T2w-STIR) and with DLR (Sag T2w-STIR-DLR). The assessment covered artifacts, perceptual signal-to-noise ratio, clearness of tissue interfaces, fat suppression, overall image quality, and the delineation of spinal cord, vertebrae, discs, dopamine, and joints. Spanning canal stenosis, neural foraminal stenosis, herniated discs, annular fissures, hypertrophy of the ligamentum flavum or vertebral facet joints, and intervertebral disc degeneration were evaluated by three impartial readers.

RESULTS: Sag T2w-STIR-DLR images exhibited markedly superior performance across quality indicators (median = 4 or 5) compared to Sag T2w-STIR sequences (median = 3 or 4) (p < 0.001). No statistically significant differences were observed between the two sequences in terms of diagnosis and grading (p > 0.05). The interobserver agreement for Sag T2w-STIR-DLR images (0.604-0.931) was higher than the other (0.545-0.853), Sag T2w-STIR-DLR (0.747-1.000) demonstrated increased concordance between reader 1 and reader 3 in comparison to Sag T2w-STIR (0.508-1.000). Acquisition time diminished from 364 to 197 s through the DLR scheme.

CONCLUSIONS: Our investigation establishes that 3.0 T fast MRI images subjected to DLR processing present heightened image quality, bolstered diagnostic performance, and reduced scanning durations for cervical spine MRI compared with conventional sequences.