DOI 10.1302/0301-620x.106b3.bjj-2023-0643.r1
Metal artifact reduction sequence-MRI abnormalities in asymptomatic patients with dual-mobility hip prostheses.
Аннотация
AIMS: Modular dual-mobility (DM) articulations are increasingly used during total hip arthroplasty (THA). However, concerns remain regarding the metal liner modularity. This study aims to correlate metal artifact reduction sequence (MARS)-MRI abnormalities with serum metal ion levels in patients with DM articulations.
METHODS
A total of 45 patients (50 hips) with a modular DM articulation were included with mean follow-up of 3.7 years (SD 1.2). Enrolled patients with an asymptomatic, primary THA and DM articulation with over two years' follow-up underwent MARS-MRI. Each patient had serum cobalt, chromium, and titanium levels drawn. Patient satisfaction, Oxford Hip Score, and Forgotten Joint Score-12 (FJS-12) were collected. Each MARS-MRI was independently reviewed by fellowship-trained musculoskeletal radiologists blinded to serum ion levels.
RESULTS
Overall, two patients (4.4%) had abnormal periprosthetic fluid collections on MARS-MRI with cobalt levels > 3.0 μg/l. Four patients (8.9%) had MARS-MRI findings consistent with greater trochanteric bursitis, all with cobalt levels < 1.0 μg/l. A seventh patient had a periprosthetic fluid collection with normal ion levels. Of the 38 patients without MARS-MRI abnormalities, 37 (97.4%) had cobalt levels < 1.0 μg/l, while one (2.6%) had a cobalt level of 1.4 μg/l. One patient (2.2%) had a chromium level > 3.0 μg/l and a periprosthetic fluid collection. Of the 41 patients with titanium levels, five (12.2%) had titanium levels > 5.0 μg/l without associated MARS-MRI abnormalities.
CONCLUSION
Periprosthetic fluid collections associated with elevated serum cobalt levels in patients with asymptomatic DM articulations occur infrequently (4.4%), but further assessment is necessary due to implant heterogeneity.