Two suspected cases of immunoglobulin-mediated interference causing falsely low vancomycin concentrations with the Beckman PETINIA method.

Gunther Mary, Saxinger Lynora, Gray Margaret, Legatt Donald

Клинический случайСтатьяПеревод в очереди

Аннотация

OBJECTIVE

To describe 2 recent cases of suspected immunoglobulin-mediated interference with the Beckman Coulter particle-enhanced turbidimetric inhibition immunoassay (PETINIA) used to measure vancomycin serum or plasma concentrations and to review the existing literature.

CASE SUMMARY: A 64-year-old woman with a history of multiple immune-related comorbidities received vancomycin for treatment of a prosthetic joint infection growing coagulase-negative Staphylococcus spp. A 33-year-old man with a history of Felty syndrome received vancomycin for the treatment of methicillin-resistant Staphylococcus aureus pneumonia. Both patients had multiple vancomycin trough concentrations determined using the Beckman Coulter PETINIA method and had measured concentrations reported as less than 4 mg/L despite appropriate vancomycin dosing for their size, age, and organ function. The patients' serum was then tested by alternative methods, which reported vancomycin concentrations consistent with those expected with the patients' dosing regimens.

DISCUSSION: Immunoglobulins are well known for interfering with chemistry assays. It is suspected that high levels of immunoglobulins in these 2 patients interfered with the accurate measurement of serum vancomycin concentrations. An objective causality assessment supported the interactions as being definite and probable, respectively. Two other cases of suspected immunoglobulin interference with vancomycin measurement have been reported in the literature when vancomycin concentrations were initially measured using the Beckman Coulter PETINIA method. In vitro studies support the hypothesis that immunoglobulin interference with vancomycin measurement by this method occurs and is a graded phenomenon.

CONCLUSIONS

These cases suggest a high likelihood of immunoglobulin-mediated interference with the Beckman Coulter PETINIA method, which results in the reporting of falsely low vancomycin serum concentrations. When vancomycin concentrations do not correlate with what would be expected clinically, clinicians should have a high index of suspicion for potential assay interference.

Для цитирования: Gunther Mary, Saxinger Lynora, Gray Margaret, Legatt Donald. Two suspected cases of immunoglobulin-mediated interference causing falsely low vancomycin concentrations with the Beckman PETINIA method.. The Annals of pharmacotherapy. 2013. DOI: 10.1345/aph.1r566.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.