Reliability of slice-encoding for metal artefact correction (SEMAC) MRI to identify prosthesis loosening in patients with painful total hip arthroplasty - a single centre, prospective, surgical validation study.
Аннотация
OBJECTIVES
To validate reliability of slice-encoding for metal artefact correction (SEMAC)-MRI findings in prosthesis loosening detection by comparing them to surgical outcomes (gold standard) in symptomatic patients following hip arthroplasties. To evaluate periprosthetic anatomical structures in symptomatic patients to identify an alternative cause of hip symptoms.
METHODS
We prospectively followed 47 symptomatic patients (55 hips, 39 painful hips - group P and 16 control hips - group C) at our institution from 2011 to 2016. We acquired 1.5 T MRI conventional and SEMAC-MRI images for all patients. Two consultants scored MRI for osteolysis and marrow oedema zone-wise using predefined signal characteristics and settled scoring variations by consensus. We used Spearman Rank-Order Correlation for correlation analysis and used OMERACT (Outcome Measures in Rheumatology) filter pillars to validate SEMAC-MRI findings.
RESULTS
Eleven patients needed revision surgery, all from group P. None from group C required revision surgery. Remaining 28 hips in the group P were managed conservatively pain completely resolved in 21 hips, eight hips had trochanteric bursitis, eight had extraarticular cause and the remaining five hips had spontaneous pain resolution. We found moderate-to-weak correlation between SEMAC-MRI findings for prosthesis loosening and revision surgery outcomes. Sensitivity, Specificity, PPV and NPV in Group P were (72.7, 64.3, 44.4, 85.7%) in T1W-SEMAC, (90.9, 46.4, 40.0, 92.9%) in STIR-SEMAC and (36.3, 78.5, 40.0, 75.8%) in PDW-SEMAC.
CONCLUSION
Negative SEMAC-MRI results can effectively exclude prosthesis loosening confirmed on revision surgery and SEMAC-MRI can detect alternative cause of hip pain accurately.
ADVANCES IN KNOWLEDGE: Negative SEMAC-MRI in painful THA patients can effectively exclude prosthesis loosening as a cause.