DOI 10.1016/j.ejmp.2022.10.012

Медицинская библиотека · Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2022

Evaluation of CBCT based dose calculation in the thorax and pelvis using two generic algorithms.

Thing R S, Nilsson R, Andersson S, Berg M, Lund M D

СтатьяПеревод в очереди

Аннотация

PURPOSE: This work examines the dosimetric performance of two algorithms creating a corrected CBCT (corrCBCT) and a virtual CT (vCT) implemented in a commercial treatment planning system.

METHODS

60 patients distributed across all patient groups treated with curative intent at Vejle Hospital (breast, lung, prostate and anal/rectal cancer) were selected for the present study. Clinical treatment plans were recalculated on corrCBCT and vCT, as well as a reference CT (refCT) acquired as close in time to the CBCT image as possible. Recalculated doses were compared using gamma analysis, as well as by comparing D98%, D50%, and D2% for all delineated targets and organs at risk.

RESULTS

High dosimetric accuracy is demonstrated on both the corrCBCT and vCT. Gamma 2%/2mm pass rates >98% were found for all patients except two outliers still having >93% pass rates. Equivalence of all evaluated dose metrics within ±1Gy was observed for all patient groups, while the pelvic patients additionally showed equivalence for all metrics within ±1% of the refCT dose. For the thoracic patients, equivalence within ±2.5% was established for all metrics except median dose to the ipsilateral lung, calculated on corrCBCT for the breast patient group.

CONCLUSION

The corrCBCT and vCT images are shown in excellent dosimetric agreement with refCT images, and show high potential for future use for streamlined adaptive radiotherapy workflows.

Для цитирования: Thing R S, Nilsson R, Andersson S, Berg M, Lund M D. Evaluation of CBCT based dose calculation in the thorax and pelvis using two generic algorithms.. Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB). 2022. DOI: 10.1016/j.ejmp.2022.10.012.