DOI 10.5114/jcb.2025.148107

Медицинская библиотека · Journal of contemporary brachytherapy · 2025

Skin marker for MR-only surface high-dose-rate brachytherapy.

Kaza Evangelia, Devlin Phillip M, O'Farrell Desmond, Buzurovic Ivan

СтатьяПеревод в очереди

Аннотация

PURPOSE: Recent advances in surface high-dose-rate (HDR) brachytherapy imaging indicate that flap applicators, human skin, and fibromatosis can be visualized using MRI. Complete MR-only surface brachytherapy workflows would require skin marker identification to define clinical target edges. However, CT markers are not detected on MR images, and common MR markers are unsuitable for continuous surface target tracing. In this paper, we proposed an alternative skin marker that was evaluated for MRI and CT detectability and contourability using a brachytherapy treatment planning system (TPS).

MATERIAL AND

METHODS

Commercially obtained silicone rubber tubes of 2 or 3 mm diameter were taped on the hand of an anthropomorphic phantom, a healthy volunteer, and three palmar fascial fibromatosis patients. Subjects were imaged with an optimized 3D pointwise encoding time reduction with radial acquisition (PETRA) sequence, and a volumetric interpolated breath-hold examination (VIBE) sequence with Dixon reconstruction. Additionally, patients underwent standard CT imaging. Obtained images were reviewed for tube conspicuity, and tubes were tracked on axial views using Oncentra Brachy TPS. Independent tube and muscle reference contours were drawn in MIM for quantitative analysis, considering the three orthogonal imaging planes.

RESULTS AND

CONCLUSIONS

Silicone rubber tubes were detected with positive signal on PETRA, VIBE, and CT images. Among the MR series, Dixon VIBE fat-only showed the highest contrast against muscle tissue and the best separation from human skin, followed by DIXON opposed-phase. 3 mm diameter tubes were tracked better by TPS than 2 mm diameter ones. Considering MR images in the three orthogonal planes in MIM was more helpful for localizing the entire tube than using axial images only in TPS. All obtained contour shapes generally agreed with the known tube positions. Overall, solid silicone rubber tubes of 3 mm diameter represent a suitable skin marker alternative to CT markers for MR-only surface HDR brachytherapy.

Для цитирования: Kaza Evangelia, Devlin Phillip M, O'Farrell Desmond, Buzurovic Ivan. Skin marker for MR-only surface high-dose-rate brachytherapy.. Journal of contemporary brachytherapy. 2025. DOI: 10.5114/jcb.2025.148107.