MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1155/2022/7123139

Медицинская библиотека · BioMed research international · 2022

Prediction of Sacral Screw Loosening after Lumbosacral Surgeries Involving Rigid Fixation of Sacral Bone Using Preoperative Computed Tomography Scans.

Muheremu Aikeremujiang, Yakufu Maihemuti, Jiang Junyao, Mardan Muradil, Li Lei, Zhang Rui, Aili Abudunaibi, Luo Zhaohui

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To find a preoperative computed tomography-based method to predict the incidence of sacral screw loosening and assist surgical planning.

METHODS

Surgically treated patients for degenerative lumbosacral disorders with rigid pedicle screw fixation of patients with L5-S1 vertebra in our center from January 2016 to January 2021 were retrospectively included in the current study. CT scan attenuation of the horizontal plane of the sacrum was measured with Hounsfield units (HU). Postoperative X-ray tests were used to diagnose screw loosening. The data was analyzed by independent sample t-tests, X 2 analysis, Pearson correlation analysis, and ROC curve analysis.

RESULTS

A total of 162 (114 male, 48 female, average age 63.7 ± 7.3 years) patients were included in the final analysis. Significant differences were found between the screw loosening group and nonloosening group concerning the HU value of the sacrum at the horizontal plane (P < 0.01). In ROC curve analysis, AUC was 0.674 (95% CI: 0.592-0.756). A cutoff of 200 HU provided 64.8% sensitivity and 62.4% specificity, and a cutoff of 150 HU provided 90.2% sensitivity.

CONCLUSIONS

Analyzing 162 patients with at least 12 months of follow-up, we propose cutoff CT attenuation values of 200 HU and 150 HU to take moderate and radical measures of screw augmentation to prevent screw loosening in the sacral bone.

Для цитирования: Muheremu Aikeremujiang, Yakufu Maihemuti, Jiang Junyao, Mardan Muradil, Li Lei, Zhang Rui, Aili Abudunaibi, Luo Zhaohui. Prediction of Sacral Screw Loosening after Lumbosacral Surgeries Involving Rigid Fixation of Sacral Bone Using Preoperative Computed Tomography Scans.. BioMed research international. 2022. DOI: 10.1155/2022/7123139.