DOI 10.1016/j.ajo.2021.06.026

Медицинская библиотека · American journal of ophthalmology · 2021

False Positive Responses in Standard Automated Perimetry.

Heijl Anders, Patella Vincent Michael, Flanagan John G, Iwase Aiko, Leung Christopher K, Tuulonen Anja, Lee Gary C, Callan Thomas, Bengtsson Boel

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

PURPOSE: To analyze the relationship between rates of false positive (FP) responses and standard automated perimetry results.

DESIGN

Prospective multicenter cross-sectional study.

METHODS

One hundred twenty-six patients with manifest or suspect glaucoma were tested with Swedish Interactive Thresholding Algorithm (SITA) Standard, SITA Fast, and SITA Faster at each of 2 visits. We calculated intervisit differences in mean deviation (MD), visual field index (VFI), and number of statistically significant test points as a function of FP rates and also as a function of general height (GH).

RESULTS

Increasing FP values were associated with higher MD values for all 3 algorithms, but the effects were small, 0.3 dB to 0.6 dB, for an increase of 10 percentage points of FP rate, and for VFI even smaller (0.6%-1.4%). Only small parts of intervisit differences were explained by FP (r2 values 0.00-0.11). The effects of FP were larger in severe glaucoma, with MD increases of 1.1 dB to 2.0 dB per 10 percentage points of FP, and r2 values ranging from 0.04 to 0.33. The numbers of significantly depressed total deviation points were affected only slightly, and pattern deviation probability maps were generally unaffected. GH was much more strongly related to perimetric outcomes than FP.

CONCLUSIONS

Across 3 different standard automated perimetry thresholding algorithms, FP rates showed only weak associations with visual field test results, except in severe glaucoma. Current recommendations regarding acceptable FP ranges may require revision. GH or other analyses may be better suited than FP rates for identifying unreliable results in patients who frequently press the response button without having perceived stimuli.

Для цитирования: Heijl Anders, Patella Vincent Michael, Flanagan John G, Iwase Aiko, Leung Christopher K, Tuulonen Anja, Lee Gary C, Callan Thomas, Bengtsson Boel. False Positive Responses in Standard Automated Perimetry.. American journal of ophthalmology. 2021. DOI: 10.1016/j.ajo.2021.06.026.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.