DOI 10.1002/mp.17325

Медицинская библиотека · Medical physics · 2024

Technical note: Phantom-based evaluation of CBCT dose calculation accuracy for use in adaptive radiotherapy.

Wessels Claas, Strzelecki Adam, Plamondon Mathieu, Lehmann Mathias, Peterlik Igor, Paysan Pascal, Nagy Balazs, Heinz Alexander, Seghers Dieter, Thompson Stephen, Scheib Stefan G

СтатьяОценочное исследованиеПеревод в очереди

Аннотация

BACKGROUND

High-quality 3D-anatomy of the day is needed for treatment plan adaptation in radiotherapy. For online x-ray-based CBCT workflows, one approach is to create a synthetic CT or to utilize a fan-beam CT with corresponding registrations. The former potentially introduces uncertainties in the dose calculation if deformable image registration is used. The latter can introduce burden and complexity to the process, the facility, and the patient.

PURPOSE: Using the CBCT of the day, acquired on the treatment device, for direct dose calculation and plan adaptation can overcome these limitations. This study aims to assess the accuracy of the calculated dose on the CBCT scans acquired on a Halcyon linear accelerator equipped with HyperSight.

METHODS

HyperSight's new CBCT reconstruction algorithm includes improvements in scatter correction, HU calibration of the imager, and beam shape adaptation. Furthermore, HyperSight introduced a new x-ray detector. To show the effect of the implemented improvements, gamma comparisons of 2%/2 mm, 2%/1 mm, and 1%/1 mm were made between the dose distribution in phantoms calculated on the CBCT reconstructions and the simulation CT scans, considering this the standard of care. The resulting gamma passing rates were compared to those obtained with the Halcyon 3.0 reconstruction and hardware without HyperSight's technologies. Various anatomical phantoms for dosimetric evaluations on brain, head and neck, lung, breast, and prostate cases have been used in this study.

RESULTS

The overall results demonstrated that HyperSight outperformed the Halcyon 3.0 version. Based on the gamma analysis, the calculated dose using HyperSight was closer to the CT scan-based doses than the calculated dose using iCBCT Halcyon 3.0 for most cases. Over all plans and gamma criteria, Halcyon 3.0 achieved an average passing rate of 92.9%, whereas HyperSight achieved 98.1%.

CONCLUSION

Using HyperSight CBCT images for direct dose calculation, for example, in (online) plan adaptation, seems feasible for the investigated cases.

Для цитирования: Wessels Claas, Strzelecki Adam, Plamondon Mathieu, Lehmann Mathias, Peterlik Igor, Paysan Pascal, Nagy Balazs, Heinz Alexander, Seghers Dieter, Thompson Stephen, Scheib Stefan G. Technical note: Phantom-based evaluation of CBCT dose calculation accuracy for use in adaptive radiotherapy.. Medical physics. 2024. DOI: 10.1002/mp.17325.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.