BACKGROUND

Evidence of any benefit in the long-term performance of highly crosslinked polyethylene (HXLPE) for total knee arthroplasty (TKA) is lacking. We sought to evaluate revision risk for HXLPE versus conventional polyethylene (CPE).

METHODS

Data from a US-based health care system's Total Joint Replacement Registry were used to conduct a cohort study. Adult patients who underwent primary fixed-bearing TKA with patella resurfacing for osteoarthritis were included (2001 to 2023). The final study sample included 232,527 TKAs: 105,028 HXLPE and 127,499 CPE. The primary outcome was revision out to 20-year follow-up, and revision for any aseptic reasons as well as for wear and loosening were the secondary outcomes. Multivariable Cox regressions were used to evaluate revision risk by polyethylene with covariate adjustment.

RESULTS

At 20-year follow-up, crude revision incidence was 6.0% for HXLPE and 6.1% for CPE. In adjusted analyses, a lower revision risk (hazard ratio [HR] = 0.93, 95% confidence interval [CI] = 0.88 to 0.98) was observed when comparing HXLPE with CPE. This lower risk was observed with other revision reasons evaluated (overall aseptic: HR = 0.93, 95% CI = 0.88 to 1.00; wear: HR = 0.75, 95% CI = 0.60 to 0.94; loosening: HR = 0.75, 95% CI = 0.67 to 0.84).

CONCLUSIONS

In this registry-based cohort study, 20-year revision incidence was approximately 6% regardless of polyethylene used. Still, we observed HXLPE to perform at least as well, if not better than CPE, at long-term follow-up in adjusted analyses, specifically for wear and loosening, supporting increased HXLPE use for TKA.

LEVEL OF EVIDENCE: Level III.

Для цитирования: Prentice Heather A, Chan Priscilla H, Fasig Brian H, Kelly Matthew P, Hinman Adrian D, Kurtz Steven M, Paxton Elizabeth W. Highly Crosslinked Polyethylene Is Associated With a Lower Revision Risk Out to 20-Year Follow-Up Versus Conventional Polyethylene in Total Knee Arthroplasty.. The Journal of arthroplasty. 2025. DOI: 10.1016/j.arth.2025.07.005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.