MedElevator Войти в исследования
← Поиск публикаций

DOI 10.22603/ssrr.2025-0094

Медицинская библиотека · Spine surgery and related research · 2025

Optimal Placement of Supplemental Accessory Rods to Prevent Rod Fracture at the Lumbosacral Junction in Long Spinopelvic Fixation Using Lateral Interbody Fusion: A Biomechanical Experimental Study Using a Synthetic Bone Model and a Finite Element Model.

Nakanishi Ryuichiro, Tsutsui Shunji, Yamamoto Ei, Kozaki Takuhei, Murata Akimasa, Yamada Hiroshi

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

High rates of postoperative rod fracture at the lumbosacral junction have been reported after long spinopelvic fixation. In the prevention of rod fractures, supplemental accessory rods (ARs) and lateral interbody fusion are commonly used and reportedly effective. However, the optimal AR placement to mitigate rod stress at the lumbosacral junction is unclear. We therefore used a synthetic bone model and a finite element model concurrently to address their respective shortcomings.

METHODS

Both models included the lumbar spine (L1-L5) and the pelvis, and were instrumented with a screw and rod system and lateral interbody fusion cages to closely resemble actual surgical procedures. The four different constructs were: two primary rods (PRs) without ARs, PRs+contoured long ARs, PRs+short ARs, and PRs+straight long ARs. In our synthetic model, we applied vertical load to the constructs and measured rod strain at L5-S1 using strain gauges. We calculated a mean value of the five rods in each construct. In our finite element model, we measured maximum principal stresses at L5-S1 after the application of flexion/extension, lateral bending, and axial rotation loads.

RESULTS

In our synthetic bone model, there was significant reduction of rod strain by 52% in PRs+straight long ARs compared with PRs without ARs (p=0.023). A reduction of average principal stress in the finite element model was observed in PRs+straight long ARs by up to 44.2% (highest against flexion load) compared with PRs without ARs.

CONCLUSIONS

We conducted concurrent biomechanical analyses using a synthetic bone model and a finite element model. We recommend straight long ARs to prevent rod fracture at the lumbosacral junction in long spinopelvic fixation.

Для цитирования: Nakanishi Ryuichiro, Tsutsui Shunji, Yamamoto Ei, Kozaki Takuhei, Murata Akimasa, Yamada Hiroshi. Optimal Placement of Supplemental Accessory Rods to Prevent Rod Fracture at the Lumbosacral Junction in Long Spinopelvic Fixation Using Lateral Interbody Fusion: A Biomechanical Experimental Study Using a Synthetic Bone Model and a Finite Element Model.. Spine surgery and related research. 2025. DOI: 10.22603/ssrr.2025-0094.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.