MedElevator Войти в исследования
← Поиск публикаций

DOI 10.7507/1002-1892.202511085

Медицинская библиотека · Zhongguo xiu fu chong jian wai ke za zhi = Zhongguo xiufu chongjian waike zazhi = Chinese journal of reparative and reconstructive surgery · 2026

[Experimental study of patient-specific three-dimensional-printed drill guide-assisted C 2 pedicle screw placement in models derived from patients with basilar invagination].

Li Jinhu, Jiang Weili, Cheng Jiawei, Niu Chong, Si Wenshuo, Niu Guoqi

Аннотация на английскомСтатьяПеревод в очереди

Аннотация

OBJECTIVE

To evaluate the screw-placement accuracy and technical advantages of patient-specific three-dimensional (3D)-printed drill guide-assisted C 2 pedicle screw placement in bone models derived from patients with basilar invagination, and to provide evidence for clinical screw-placement strategies in high-risk craniocervical junction deformities.

METHODS

Preoperative thin-slice CT data from 9 patients with basilar invagination treated between June 2016 and July 2025 were used. For each patient, two 1∶1-scale 3D-printed upper cervical spine models were fabricated, and assigned to the guide-assisted group and the freehand group by computer-generated block randomization. Bilateral C 2 screws were inserted by the same spine surgeon according to a standardized protocol. In the guide-assisted group, screw placement was performed in the following sequence: guide positioning, guide pin insertion, guide removal followed by enlargement drilling, tapping, and screw insertion. In the freehand group, screw placement was performed by anatomical landmark-based localization, stepwise drilling with fluoroscopic correction, enlargement drilling and tapping, and screw insertion. The primary outcomes were postoperative CT-based Gertzbein-Robbins (GR) grading of screw position and the GR grade 0/1 acceptability rate. Secondary outcomes included entry-point deviation, axial angle deviation, and the number of fluoroscopic exposures.

RESULTS

The two assessors showed good agreement in GR grading, with a Kappa coefficient>0.80, indicating reliable evaluations. Postoperative CT assessment showed that the overall distribution of GR grades was significantly better in the guide-assisted group than in the freehand group ( P<0.05). The GR grade 0/1 acceptability rate was higher in the guide-assisted group, but the difference was not significant ( P>0.05). Entry-point deviation, axial angle deviation, and the number of fluoroscopic exposures were all significantly lower in the guide-assisted group than in the freehand group ( P<0.05).

CONCLUSION

Patient-specific 3D-printed drill guides can reduce geometric deviations in C 2 pedicle screw placement in basilar invagination and decrease the number of fluoroscopic exposures. They show a potential advantage in improving the GR grade 0/1 acceptability rate; however, further clinical studies with larger samples are needed for verification.

Для цитирования: Li Jinhu, Jiang Weili, Cheng Jiawei, Niu Chong, Si Wenshuo, Niu Guoqi. [Experimental study of patient-specific three-dimensional-printed drill guide-assisted C 2 pedicle screw placement in models derived from patients with basilar invagination].. Zhongguo xiu fu chong jian wai ke za zhi = Zhongguo xiufu chongjian waike zazhi = Chinese journal of reparative and reconstructive surgery. 2026. DOI: 10.7507/1002-1892.202511085.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.