BACKGROUND
Electromechanical reshaping is a novel, minimally invasive means to induce mechanical changes in connective tissues, and has the potential to be utilized in lieu of current orthopedic therapies that involve tendons and ligaments. Electromechanical reshaping delivers an electrical current to tissues while under mechanical deformation, causing in situ redox changes that produce reliably controlled and spatially limited mechanical and structural changes. In this study, we examine the feasibility of altering Young's modulus and inducing a shape deformation using an ex vivo bovine Achilles tendon model.
METHODS
Tendon was mechanically deformed in two different modes: (1) elongation to assess for tensile modulus and (2) compression to assess for compressive modulus. Electromechanical reshaping was applied to tendon specimens via flat plate platinum electrodes (6 V, 3 min) while simultaneously under mechanical strain for 15 min.
FINDINGS: In elongation mode, post-electromechanical reshaping samples demonstrated a significant decrease in Young's modulus compared to pretreatment samples (66.02 and 45.12 MPa, respectively, p < 0.0049). In compression mode, posttreatment samples illustrated a significant shape change, with an increase in diameter (10.62 to 11.36 mm, p < 0.05) and decrease in thickness (4.13 to 3.62 mm, p < 0.05).
INTERPRETATION: Results demonstrated a tissue softening effect without lengthening deformation during elongation, and a shortening effect without compromising compressive stiffness during compression. Electromechanical reshaping's reliable, low-cost, and efficacious methodology in inducing mechanical and structural connective tissue modifications illustrates a potential for future alternative orthopedic applications. Future studies will optimize and refine electromechanical reshaping to address clinically relevant geometries and methods such as needle techniques.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.