Dual-Plate Allograft Reconstruction for Expansile Lytic Lesions of the Metacarpal May Allow for Immediate Postoperative Weightbearing in a Biomechanical Setting.

Sridharan Mathangi, Tomkinson Alyssa, Mefford Mikayla, Rahmati Kasra, Clites Tyler, Wessel Lauren E

СтатьяПеревод в очереди

Аннотация

BACKGROUND

This biomechanical study investigates whether reconstruction of expansile lesions of the metacarpal with fibular allograft and a dual-plate construct is strong enough for immediate postoperative weightbearing.

METHODS

Twelve cadaveric middle finger metacarpals (6 matched pairs) from adult donors were used. One side of each pair was randomly assigned to undergo allograft reconstruction with dual plating (experimental construct). The matched contralateral side was assigned to the control construct, a transverse fracture with standard, 8-hole dorsal plating. In the experimental construct, an expansile bone lesion was simulated in the native metacarpal by resecting 1.5 cm from the proximal carpometacarpal joint to 1.5 cm from the distal articular surface and was reconstructed with fibular allograft. Biomechanical testing was performed by a 6-degree-of-freedom robotic KUKA manipulator. The distal portion of each specimen was clamped in place, and a cantilever bending force was applied by the KUKA to test the load to failure. Failure was defined as periprosthetic fracture, fracture displacement, or screw pullout. Paired t tests were used to compare load to failure between matched pairs (R studio). Significance was set at P < .05.

RESULTS

The load to failure of dual-plate allograft constructs did not significantly differ from that of the control construct, 494.7 ± 239.07 N (95% CI, 243.81-745.59 N) versus 342.2 ± 99.9 N (95% CI, 237.45-447.04 N), P = .19. Experimental constructs failed at the proximal allograft-native bone interface. Control constructs failed via fracture displacement.

CONCLUSIONS

Dual-plate fixation constructs for allograft reconstruction of the metacarpal are comparable in mechanical stability to standard metacarpal fracture fixation that allow for immediate postoperative range of motion.

Для цитирования: Sridharan Mathangi, Tomkinson Alyssa, Mefford Mikayla, Rahmati Kasra, Clites Tyler, Wessel Lauren E. Dual-Plate Allograft Reconstruction for Expansile Lytic Lesions of the Metacarpal May Allow for Immediate Postoperative Weightbearing in a Biomechanical Setting.. Hand (New York, N.Y.). 2026. DOI: 10.1177/15589447261479354.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.